Comparison of inductively coupled mass spectrometry and ultra low-level gamma-ray spectroscopy for ultra low background material selection

Appl Radiat Isot. 2009 May;67(5):828-32. doi: 10.1016/j.apradiso.2009.01.021. Epub 2009 Jan 21.

Abstract

The main target of this study is the development and the validation of an internal procedure to measure the pg g(-1) levels, of naturally occurring radioisotopes to select ultra low background detector materials. As there are no ultra low-level certified reference materials commercially available, the performance of ICP-MS was compared with the well established ultra low-level gamma-spectroscopy. This work shows also the reliability of very quick ICP-MS semi-quantitative analyses of an environmental matrix such as a geological sample treated according to conventional protocols.