. 2009 Jun;18(3):213-6.
doi: 10.1136/qshc.2007.024562.

Ambulance stretcher adverse events

Affiliations

Ambulance stretcher adverse events

H E Wang et al. Qual Saf Health Care. 2009 Jun.

Abstract

Introduction: Ambulance personnel use wheeled stretchers for moving patients in the out-of-hospital setting. The nature of adverse events and associated injuries occurring during ambulance stretcher operation was characterised.

Methods: Data from the United States Food and Drug Administration's Manufacturer and User Facility Device Experience Database (MAUDE) were used. All adverse events involving ambulance stretchers during the years 1996-2005 were identified. The nature of the event, the method of stretcher handling, the individuals injured and the nature of the resulting injuries were identified.

Results: There were 671 reported adverse events. The most common adverse events were stretcher collapse (54%; 95% CI 50 to 57%), broken, missing or malfunctioning part (28%; 95% CI 25 to 32%) and dropped stretcher (7%; 95% CI 5 to 9%). Adverse events most commonly occurred during unloading of the stretcher from the ambulance (16%; 13 to 19%). Injuries occurred in 121 events (18%; 95% CI 15 to 21%), most often involving sprains/strains (29%), fractures (16%) and lacerations/avulsions (13%). There were three traumatic brain injuries and three deaths. Patients sustained injuries in 52 events (43%), and ambulance personnel sustained injuries in 64 events (53%). More than one individual sustained injuries in 12 events.

Conclusion: Adverse events may occur during ambulance stretcher operation and can result in significant injury to patients and ambulance personnel.