Shaping ability of four nickel-titanium rotary instruments in simulated S-shaped canals

J Endod. 2009 Jun;35(6):883-6. doi: 10.1016/j.joen.2009.03.007. Epub 2009 Apr 22.

Abstract

Introduction: The purpose of this study was to compare the shaping ability of ProTaper, Mtwo, BioRaCe, and BioRaCe + S-Apex instruments in simulated canals with an S-shaped curvature.

Methods: Canal transportation and aberrations were assessed by comparing the preinstrumentation and postinstrumentation images under a stereomicroscope. Analysis of variance and post hoc Student-Newman-Keuls test were used for statistical analysis.

Results: ProTaper instruments caused more pronounced canal transportation in the apical curvature (P < .01) than all other instruments. The use of ProTaper, Mtwo, and BioRaCe instruments resulted in more canal aberrations compared with BioRaCe + S-Apex (P < .05).

Conclusions: NiTi systems including less tapered and more flexible instruments like S-Apex seem to be favorable when preparing S-shaped canals.

MeSH terms

  • Dental Alloys
  • Dental Instruments*
  • Dental Pulp Cavity / anatomy & histology*
  • Equipment Design
  • Equipment Failure
  • Humans
  • Models, Dental
  • Nickel
  • Root Canal Preparation / instrumentation*
  • Titanium
  • Tooth Apex / anatomy & histology

Substances

  • Dental Alloys
  • titanium nickelide
  • Nickel
  • Titanium