Acute evaluation of the Mynx vascular closure device during arterial re-puncture in an ovine model

J Invasive Cardiol. 2009 Jun;21(6):283-5.

Abstract

Background: The Mynx vascular closure device (VCD) (AccessClosure, Inc., Mountain View, California) is used for extravascular closure of the common femoral artery (CFA) after diagnostic or interventional endovascular procedures. Although it is common clinical practice to re-puncture a CFA, for some VCDs, acute repuncture is discouraged, while for other VCDs the safety of direct re-puncture through a previously placed VCD is still in question. The objective of this preclinical study was to evaluate the acute performance of the Mynx device where direct re-puncture has occurred through the hydrogel sealant, followed by delivery of a second Mynx device to close the reaccess puncture site.

Methods: Two ovine models with a total of 24 punctures (12 sets of puncture and re-puncture) were used for this study. Success was measured utilizing angiographic and observational endpoints.

Results: There was no evidence of hydrogel sealant prolapse into the artery, distal embolization of sealant following repuncture, groin site bleeding or hematoma. Successful hemostasis was achieved in all punctures (mean activated clotting time = 357 secs) with the exception of a minor (non-clinically significant) contrast extravasation in the tissue tract, which resolved with 2 minutes of adjunctive compression.

Conclusion: Arterial re-puncture in an ovine model directly through a recently placed Mynx closure device and subsequent re-closure with a second Mynx device can be successfully accomplished.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Embolism / prevention & control
  • Equipment and Supplies* / adverse effects
  • Femoral Artery / surgery*
  • Hematoma / prevention & control
  • Hemorrhage / prevention & control
  • Hemostasis, Surgical / instrumentation*
  • Hemostasis, Surgical / methods
  • Hydrogel, Polyethylene Glycol Dimethacrylate* / adverse effects
  • Male
  • Models, Animal
  • Sheep
  • Vascular Surgical Procedures / adverse effects
  • Vascular Surgical Procedures / instrumentation*
  • Vascular Surgical Procedures / methods

Substances

  • Hydrogel, Polyethylene Glycol Dimethacrylate