Background and purpose: Dual-energy X-ray absorptiometry (DXA) analysis of the 7 periprosthetic Gruen zones is the most commonly used protocol to evaluate bone remodeling after the implantation of conventional femoral stems. We assessed the value of DXA after cementless primary total hip arthroplasty (THA) by comparing the effect of progressive shortening of the stem of two femoral implants on periprosthetic bone remodeling using a specifically developed protocol of analysis with 5 periprosthetic regions of interest (ROIs).
Patients and methods: Bone mineral density (BMD) was evaluated in 37 patients in the plateau stage, 3 years after THA. Two femoral implants featuring conceptually new designs and surgical technique were tested: types 1 and 2, characterized by extremely short stem and virtual absence of distal stem, respectively.
Results: We found that progressive shortening of the femoral stem produces more proximal loading, which effectively preserves metaphyseal bone stock and increases periprosthetic BMD in the medial ROIs over time. In the type 2 group, higher absolute BMD values were observed in medial ROIs 4 and 5. No differences were found in ROIs 1, 2, and 3.
Interpretation: This study shows the flexibility of DXA in adapting the protocol of periprosthetic analysis to the specific requirements of new implant designs, and it shows its high sensitivity in evaluation of the biological response of bone to changes in implant shape.