Assessment of artifacts and reproducibility across spectral- and time-domain optical coherence tomography devices

Ophthalmology. 2009 Oct;116(10):1960-70. doi: 10.1016/j.ophtha.2009.03.034. Epub 2009 Jul 9.

Abstract

Purpose: To report the frequency of optical coherence tomography (OCT) scan artifacts and to compare macular thickness measurements, interscan reproducibility, and interdevice agreeability across 3 spectral-domain (SD) OCT (also known as Fourier domain; Cirrus HD-OCT, RTVue-100, and Topcon 3D-OCT 1000) devices and 1 time-domain (TD) OCT (Stratus OCT) device.

Design: Prospective, noncomparative, noninterventional case series.

Participants: Fifty-two patients seen at the New England Eye Center, Tufts Medical Center Retina Service, between February and August 2008.

Methods: Two scans were performed for each of the SD OCT protocols: Cirrus macular cube 512 x 128 (software version 3.0; Carl Zeiss Meditec, Inc., Dublin, CA), RTVue (E)MM5 and MM6 (software version 3.5; Optovue, Inc., Fremont, CA), Topcon 3D Macular and Radial (software version 2.12; Topcon, Inc., Paramus, NJ), in addition to 1 TD OCT scan via Stratus macular thickness protocol (software version 4.0; Carl Zeiss Meditec, Inc.). Scans were inspected for 6 types of OCT scan artifacts and were analyzed. Interscan reproducibility and interdevice agreeability were assessed by intraclass correlation coefficients (ICCs) and Bland-Altman plots, respectively.

Main outcome measures: Optical coherence tomography image artifacts, macular thickness, reproducibility, and agreeability.

Results: Time-domain OCT scans contained a significantly higher percentage of clinically significant improper central foveal thickness (IFT) after manual correction (11-mum change or more) compared with SD OCT scans. Cirrus HD-OCT had a significantly lower percentage of clinically significant IFT (11.1%) compared with the other SD OCT devices (Topcon 3D, 20.4%; Topcon Radial, 29.6%; RTVue (E)MM5, 42.6%; RTVue MM6, 24.1%; P = 0.001). All 3 SD OCT devices had central foveal subfield thicknesses that were significantly more than that of TD OCT after manual correction (P<0.0001). All 3 SD OCT devices demonstrated a high degree of reproducibility in the central foveal region (ICCs, 0.92-0.97). Bland-Altman plots showed low agreeability between TD and SD OCT scans.

Conclusions: Out of all OCT devices analyzed, cirrus HD-OCT scans exhibited the lowest occurrence of any artifacts (68.5%), IFT (40.7%), and clinically significant IFT (11.1%), whereas Stratus OCT scans exhibited the highest occurrence of clinically significant IFT. Further work on improving segmentation algorithm to decrease artifacts is warranted.

Publication types

  • Comparative Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Artifacts*
  • Diagnostic Errors
  • Diagnostic Techniques, Ophthalmological / instrumentation*
  • Female
  • Fourier Analysis
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Reproducibility of Results
  • Retina / pathology*
  • Retinal Diseases / diagnosis*
  • Time Factors
  • Tomography, Optical Coherence*