Reference values for left ventricular volumes with real-time 3-dimensional echocardiography

Scand Cardiovasc J. 2010 Feb;44(1):24-30. doi: 10.3109/14017430903114446.

Abstract

Objectives: Obtain normal reference ranges for left ventricular (LV) volume indexes (VI) and ejection fraction (EF) with fast real-time 3-dimensional echocardiography (RT3DE) with online analysis.

Design: After a screening visit 166 healthy participants, 79 males and 87 females aged 29-80 years were examined with RT3DE and Doppler.

Results: Upper normal values (mean + 2 standard deviations [SD]) for LV end-diastolic volume index (LVEDVI) and LV end-systolic volume index (LVESVI) were 82 ml/m(2) and 38 ml/m(2), respectively. The lower limit (mean - 2 SD) for LVEF was 49%. LVVI were significantly larger among males (p < 0.001). LV stroke volume (SD) was 87 (22) ml with Doppler vs. 69 (14) ml with RT3DE (p < 0.001). In males there was a positive correlation between age and LVEF (r = 0.34, p = 0.003). In a reproducibility study of RT3DE and 2-dimensional echocardiography (2DE) the coefficients of variability for repeated recordings were 5.7% for LVEDV, 7.7% for LVESV and 6.7% for LVEF with RT3DE, and 8.6%, 8.6% and 6.4% with 2DE, respectively.

Conclusions: These reference values presented from a large group of normal subjects over a wide age range with RT3DE may represent a valuable tool to evaluate if LV systolic dysfunction is present or not.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Age Factors
  • Aged
  • Aged, 80 and over
  • Body Surface Area
  • Echocardiography, Doppler
  • Echocardiography, Three-Dimensional*
  • Female
  • Heart Ventricles / diagnostic imaging*
  • Humans
  • Image Interpretation, Computer-Assisted
  • Male
  • Middle Aged
  • Norway
  • Predictive Value of Tests
  • Reference Values
  • Reproducibility of Results
  • Sex Factors
  • Stroke Volume*
  • Ventricular Function, Left*