Numerical simulation of the effect of time-to-loading on peri-implant bone

Med Eng Phys. 2010 Jan;32(1):7-13. doi: 10.1016/j.medengphy.2009.10.001. Epub 2009 Oct 27.

Abstract

Purpose: To evaluate the effect of time-to-loading on trabecular bone around single-tooth dental implants using numerical solutions based on computer models.

Materials and methods: A global model with a coarse mesh carrying a Straumann dental implant (043.033S; Institut Straumann, Basel, Switzerland) was created. A region of interest in trabecular bone was defined to study a localized part of the global model with a refined mesh. Time-to-loading submodels to simulate 2h, 4 days, 1, 4, 6 and 12 wks of trabecular bone-healing status were designed and created. Bone types were considered in the simulation by different elastic bone properties. A 100-N oblique static load was applied. Maximum and minimum principal stresses were calculated and visualized.

Results: Bone types with higher elastic moduli experienced higher stress levels. Changes in the quality and quantity of bone at the bone-implant interface did not affect the overall stress distribution. Peri-implant bone with a higher elastic modulus preserved the stress increase at the implant-bone interface.

Discussion: Reduced bone contact may not have a prevailing effect over bone quality and quantity on stress generation at the peri-implant bone.

Conclusion: Time-to-loading of single-tooth implants may not differ in terms of load distributions in neighboring peri-implant bone.

MeSH terms

  • Biomechanical Phenomena
  • Biomedical Engineering
  • Bone and Bones / physiology*
  • Compressive Strength
  • Computer Simulation
  • Dental Implants*
  • Dental Prosthesis Design
  • Dental Prosthesis, Implant-Supported
  • Dental Stress Analysis
  • Elastic Modulus
  • Elasticity
  • Finite Element Analysis
  • Humans
  • Models, Theoretical
  • Stress, Mechanical

Substances

  • Dental Implants