Vacuum decay container/closure integrity testing technology. Part 2. Comparison to dye ingress tests

PDA J Pharm Sci Technol. 2009 Sep-Oct;63(5):489-98.

Abstract

Part 1 of this series demonstrated that a container closure integrity test performed according to ASTM F2338-09 Standard Test Method for Nondestructive Detection of Leaks in Packages by Vacuum Decay Method using a VeriPac 325/LV vacuum decay leak tester by Packaging Technologies & Inspection, LLC (PTI) is capable of detecting leaks > or = 5.0 microm (nominal diameter) in rigid, nonporous package systems, such as prefilled glass syringes. The current study compared USP, Ph.Eur. and ISO dye ingress integrity test methods to PTI's vacuum decay technology for the detection of these same 5-, 10-, and 15-microm laser-drilled hole defects in 1-mL glass prefilled syringes. The study was performed at three test sites using several inspectors and a variety of inspection conditions. No standard dye ingress method was found to reliably identify all holed syringes. Modifications to these standard dye tests' challenge conditions increased the potential for dye ingress, and adjustments to the visual inspection environment improved dye ingress detection. However, the risk of false positive test results with dye ingress tests remained. In contrast, the nondestructive vacuum decay leak test method reliably identified syringes with holes > or = 5.0 microm.

Publication types

  • Comparative Study

MeSH terms

  • Coloring Agents / chemistry*
  • Drug Contamination / prevention & control
  • Drug Industry / methods
  • Drug Packaging / standards*
  • Product Packaging / standards
  • Syringes
  • Technology, Pharmaceutical / methods*
  • Vacuum