Miniaturised centrifugal solid phase extraction platforms for in-field sampling, pre-concentration and spectrometric detection of organic pollutants in aqueous samples

Talanta. 2010 Apr 15;81(1-2):722-6. doi: 10.1016/j.talanta.2009.12.001. Epub 2009 Dec 16.

Abstract

Great variations in pollutant concentrations are observed in the environment and pre-concentration is often required to detect trace contaminants in water samples. This paper presents a novel solid phase-extraction device integrated onto a centrifugal microfluidic platform for rapid on-site pre-concentration and screening of organic contaminants in aqueous samples. In-column fluorescence and absorbance measurements are obtained directly from an analyte trapped on the top of a solid phase extraction microcolumn. Results are presented for the representative fluorophore fluorescein and the polycyclic aromatic hydrocarbon anthracene. An absolute detection limit of 20 ng was obtained for anthracene using a simple light emitting diode for fluorescence excitation. One of the main advantages of this device is that only a simple motor is needed to induce liquid flow, making simultaneous on-site extraction and measurement of multiple samples easy while minimizing sample losses and contamination.

Publication types

  • Research Support, Non-U.S. Gov't