Effect of disease severity on the performance of Cirrus spectral-domain OCT for glaucoma diagnosis

Invest Ophthalmol Vis Sci. 2010 Aug;51(8):4104-9. doi: 10.1167/iovs.09-4716. Epub 2010 Mar 24.

Abstract

Purpose: To evaluate the effect of disease severity on the diagnostic accuracy of the Cirrus Optical Coherence Tomograph (Cirrus HD-OCT; Carl Zeiss Meditec, Inc., Dublin, CA) for glaucoma detection.

Methods: One hundred thirty-five glaucomatous eyes of 99 patients and 79 normal eyes of 47 control subjects were recruited from the longitudinal Diagnostic Innovations in Glaucoma Study (DIGS). The severity of the disease was graded based on the visual field index (VFI) from standard automated perimetry. Imaging of the retinal nerve fiber layer (RNFL) was obtained using the optic disc cube protocol available on the Cirrus HD-OCT. Pooled receiver operating characteristic (ROC) curves were initially obtained for each parameter of the Cirrus HD-OCT. The effect of disease severity on diagnostic performance was evaluated by fitting an ROC regression model, with VFI used as a covariate, and calculating the area under the ROC curve (AUCs) for different levels of disease severity.

Results: The largest pooled AUCs were for average thickness (0.892), inferior quadrant thickness (0.881), and superior quadrant thickness (0.874). Disease severity had a significant influence on the detection of glaucoma. For the average RNFL thickness parameter, AUCs were 0.962, 0.932, 0.886, and 0.822 for VFIs of 70%, 80%, 90%, and 100%, respectively.

Conclusions: Disease severity had a significant effect on the diagnostic performance of the Cirrus HD-OCT and thus should be considered when interpreting results from this device and when considering the potential applications of this instrument for diagnosing glaucoma in the various clinical settings.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Area Under Curve
  • Axons / pathology*
  • Case-Control Studies
  • Female
  • Glaucoma, Open-Angle / diagnosis*
  • Gonioscopy
  • Humans
  • Male
  • Middle Aged
  • Optic Disk / pathology*
  • ROC Curve
  • Reproducibility of Results
  • Retinal Ganglion Cells / pathology*
  • Sensitivity and Specificity
  • Severity of Illness Index*
  • Tomography, Optical Coherence / instrumentation
  • Tomography, Optical Coherence / standards*
  • Visual Field Tests
  • Visual Fields