Measurements of the exerted pressure by pelvic circumferential compression devices

Open Orthop J. 2010 Feb 17;4:101-6. doi: 10.2174/1874325001004020101.

Abstract

Background: Data on the efficacy and safety of non-invasive Pelvic Circumferential Compression Devices (PCCDs) is limited. Tissue damage may occur if a continuous pressure on the skin exceeding 9.3 kPa is sustained for more than two or three hours. The aim of this study was to gain insight into the pressure build-up at the interface, by measuring the PCCD-induced pressure when applying pulling forces to three different PCCDs (Pelvic Binder((R)) , SAM-Sling ((R)) and T-POD((R)) ) in a simplified model.

Methods: The resulting exerted pressures were measured at four 'anatomical' locations (right, left, posterior and anterior) in a model using a pressure measurement system consisting of pressure cuffs.

Results: The exerted pressure varied substantially between the locations as well as between the PCCDs. Maximum pressures ranged from 18.9-23.3 kPa and from 19.2-27.5 kPa at the right location and left location, respectively. Pressures at the posterior location stayed below 18 kPa. At the anterior location pressures varied markedly between the different PCCDs.

Conclusion: The circumferential compression by the different PCCDs showed high pressures measured at the four locations using a simplified model. Difference in design and functional characteristics of the PCCDs resulted in different pressure build-up at the four locations. When following the manufacturer's instructions, the exerted pressure of all three PCCDs tested exceeded the tissue damaging level (9.3 kPa). In case of prolonged use in a clinical situation this might put patients at risk for developing tissue damage.

Keywords: PCCD; Pelvic circumferential compression device; non-invasive; pressure measurement..