The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
- PMID: 20427310
- PMCID: PMC2975961
- DOI: 10.1136/qshc.2009.032839
The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
Abstract
Objective: Assess the impact of infusion pump technologies (traditional pump vs smart pump vs smart pump with barcode) on nurses' ability to safely administer intravenous medications.
Design: Experimental study with a repeated measures design.
Setting: High-fidelity simulated inpatient unit.
Results: The nurses remedied 60% of "wrong drug" errors. This rate did not vary as a function of pump type. The nurses remedied "wrong patient" errors more often when using the barcode pump (88%) than when using the traditional pump (46%) or the smart pump (58%) (Cochran Q=14.36; p<0.05). The number of nurses who remedied "wrong dose hard limit" errors was higher when using the smart pump (75%) and the barcode pump (79%) than when using the traditional pump (38%) (Cochran Q=12.13; p<0.003). Conversely, there was no difference in remediation of "wrong dose soft limit" errors across pump types. The nurses' pump programming was less accurate when mathematical conversions were required. Success rates on secondary infusions were low (55.6%) and did not vary as a function of pump type.
Conclusions: These findings indicate that soft (changeable) limits in smart infusion pumps had no significant effect in preventing dosing errors. Provided that smart pumps are programmed with hard (unchangeable) limits, they can prevent dosing errors, thereby increasing patient safety. Until barcode pumps are integrated with other systems within the medication administration process, their role in enhancing patient safety will be limited. Further improvements to pump technologies are needed to mitigate risks associated with intravenous infusions, particularly secondary infusions.
Conflict of interest statement
Figures
Similar articles
-
Benefits and risks of using smart pumps to reduce medication error rates: a systematic review.Drug Saf. 2014 Dec;37(12):1011-20. doi: 10.1007/s40264-014-0232-1. Drug Saf. 2014. PMID: 25294653 Review.
-
Developing a drug library for smart pumps in a pediatric intensive care unit.Artif Intell Med. 2012 Mar;54(3):155-61. doi: 10.1016/j.artmed.2011.12.001. Epub 2012 Jan 4. Artif Intell Med. 2012. PMID: 22226048
-
Analysis of event logs from syringe pumps: a retrospective pilot study to assess possible effects of syringe pumps on safety in a university hospital critical care unit in Germany.Drug Saf. 2012 Jul 1;35(7):563-74. doi: 10.2165/11597350-000000000-00000. Drug Saf. 2012. PMID: 22671864
-
Nurses' risk without using smart pumps.JONAS Healthc Law Ethics Regul. 2011 Jan-Mar;13(1):17-20. doi: 10.1097/NHL.0b013e31820e0eb5. JONAS Healthc Law Ethics Regul. 2011. PMID: 21343744 Review.
-
Dose error reduction software in medication safety risk management - optimising the smart infusion pump dosing limits in neonatal intensive care unit prior to implementation.BMC Pediatr. 2022 Mar 8;22(1):118. doi: 10.1186/s12887-022-03183-8. BMC Pediatr. 2022. PMID: 35255846 Free PMC article.
Cited by
-
Detecting Unusual Intravenous Infusion Alerting Patterns with Machine Learning Algorithms.Biomed Instrum Technol. 2022 Apr 1;56(2):58-70. doi: 10.2345/0899-8205-56.2.58. Biomed Instrum Technol. 2022. PMID: 35749264 Free PMC article.
-
Quantifying the Impact of Infusion Alerts and Alarms on Nursing Workflows: A Retrospective Analysis.Appl Clin Inform. 2021 May;12(3):528-538. doi: 10.1055/s-0041-1730031. Epub 2021 Jun 30. Appl Clin Inform. 2021. PMID: 34192773 Free PMC article.
-
Integrating and Evaluating the Data Quality and Utility of Smart Pump Information in Detecting Medication Administration Errors: Evaluation Study.JMIR Med Inform. 2020 Sep 2;8(9):e19774. doi: 10.2196/19774. JMIR Med Inform. 2020. PMID: 32876578 Free PMC article.
-
Human-Based Errors Involving Smart Infusion Pumps: A Catalog of Error Types and Prevention Strategies.Drug Saf. 2020 Nov;43(11):1073-1087. doi: 10.1007/s40264-020-00986-5. Drug Saf. 2020. PMID: 32797355 Free PMC article. Review.
-
Twelve-Month Review of Infusion Pump Near-Miss Medication and Dose Selection Errors and User-Initiated "Good Save" Corrections: Retrospective Study.JMIR Hum Factors. 2020 Aug 11;7(3):e20364. doi: 10.2196/20364. JMIR Hum Factors. 2020. PMID: 32667895 Free PMC article.
References
-
- Leape LL, Brennan TA, Laird NM, et al. The nature of adverse events in hospitalized patients: results from the Harvard Medical Practice Study II. N Engl J Med 1991;324:377–84 - PubMed
-
- Kinnealey E, Fishman G, Sims N, et al. Infusion pumps with “drug libraries” at the point of care—a common solution for safer drug delivery 2003. http://www.npsf.org/download/Kinnealy.pdf (accessed 22 Dec 2008).
-
- Kaushal R, Bates DW, Landrigan C, et al. Medication errors and adverse drug events in pediatric inpatients. JAMA 2001;285:2114–20 - PubMed
-
- Tourville J. Automation and error reduction: how technology is helping Children's Medical Center of Dallas reach zero-error tolerance. US Pharmacist 2003;28:80–6
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources