Pre-analytical in-vitro stability of [-2]proPSA in blood and serum

Clin Biochem. 2010 Jul;43(10-11):926-8. doi: 10.1016/j.clinbiochem.2010.04.062. Epub 2010 May 8.

Abstract

Objectives: [-2]proPSA may discriminate prostate cancer from benign biopsy results. We characterized the pre-analytical stability of [-2]proPSA.

Design and methods: 22 volunteers, total PSA of 4.5-19.3microg/L, had blood drawn simultaneously. Baseline measurements were performed and samples were stored under various conditions prior to measurements. Freeze-thaw cycles were performed. [-2]proPSA was measured with the p2PSA automated research use only immunoassay on the Access analyzer.

Results: Mean [-2]proPSA increases with clotting time, exceeding 10% change in recovery after 3h. In serum, [-2]proPSA values decline over time under investigated storage conditions. Serum samples kept frozen show less than 10% variation in recoveries over the course of 2 freeze-thaw cycles.

Conclusions: For proper measurement of [-2]proPSA, blood samples should be centrifuged within 3h of blood draw. Serum may be stored at RT or refrigerated (+4 degrees C) for a maximum of 48h and should be frozen if stored for a longer period. Two freeze-thaw cycles have no effect on [-2]proPSA stability.