Lead(II) ion selective electrodes with PVC membranes based on two bis-thioureas as ionophores: 1,3-bis(N'-benzoylthioureido)benzene and 1,3-bis(N'-furoylthioureido)benzene

J Hazard Mater. 2010 Sep 15;181(1-3):140-6. doi: 10.1016/j.jhazmat.2010.04.107. Epub 2010 May 2.

Abstract

Two PVC membrane ion selective electrodes for Pb(II) ion based on two bis-thioureas: 1,3-bis(N'-benzoylthioureido)benzene and 1,3-bis(N'-furoylthioureido)benzene as ionophores, are reported. A first membrane formulated using 1,3-bis(N'-benzoylthioureido)benzene as carrier exhibited a Nernstian response to Pb(II) over a wide concentration range (4.0x10(-6) to 1.0x10(-2)M) with a slope of 31.5+/-1.6 mV/dec. It showed a fast response time (t(90%)=14 s) and could be used for 10 weeks without any divergence in potentials. The membrane formulated using 1,3-bis(N'-furoylthioureido)benzene as carrier exhibited a Nernstian response in the concentration range (5.0x10(-6) to 1.0x10(-2) M), with a slope of 30.0+/-1.3 mV/dec. Its response time was t(90%)=14 s, and it could be used for 14 weeks without any divergence in potentials. The two proposed potentiometric sensors revealed acceptable selectivities for Pb(II) over a wide variety of other metal ions and could be used in a pH range of 2.2-6.0. Both electrodes were assayed in direct potentiometric determination of lead in soils (10-30 mg/kg range) with very good performance (0.99935 correlation coefficient in the comparison against ICP-MS method).

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Benzene Derivatives
  • Ion-Selective Electrodes / standards*
  • Ionophores / chemistry*
  • Lead / analysis*
  • Membranes, Artificial
  • Polyvinyl Chloride
  • Soil Pollutants / analysis
  • Thiourea*

Substances

  • Benzene Derivatives
  • Ionophores
  • Membranes, Artificial
  • Soil Pollutants
  • Lead
  • Polyvinyl Chloride
  • Thiourea