A method for quantifying cellular uptake of fluorescently labeled siRNA

J Control Release. 2010 Nov 20;148(1):106-109. doi: 10.1016/j.jconrel.2010.06.019. Epub 2010 Jun 30.

Abstract

Efficient intracellular delivery of siRNA is a significant hurdle to its therapeutic success. For biological studies on the efficiency of carrier-mediated uptake of siRNA, quantitative determination of the amount of internalized siRNA is required. In this study, when the apparent uptake of fluorescently labeled siRNA, formulated in different lipo- and polyplexes, was examined using different techniques, major differences were observed. Additional experiments showed that these differences could be explained by quenching phenomena that were dependent on interactions between siRNA and carrier and their intracellular environment. Differences in fluorescent signal of complexed siRNA due to quenching could be overcome by measuring the fluorescent signal after lysing the transfected cells in lysis buffer that contained 2% SDS to dissociate siRNA from the complexes. This method offers a simple approach for quantifying cellular uptake of siRNA, which might help in the development of more efficient delivery systems.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cell Line, Tumor
  • Fluorescent Dyes
  • Gene Transfer Techniques*
  • Humans
  • Microscopy, Fluorescence
  • RNA, Small Interfering / administration & dosage
  • RNA, Small Interfering / pharmacokinetics*

Substances

  • Fluorescent Dyes
  • RNA, Small Interfering