Balanced SSFP transient imaging using variable flip angles for a predefined signal profile

Magn Reson Med. 2010 Nov;64(5):1404-12. doi: 10.1002/mrm.22541.

Abstract

Variable flip angles are used in steady-state free precession (SSFP) acquisitions (e.g., time-of-flight) but to a lesser extent than in spin echo acquisitions. In balanced steady-state free precession, imaging is often assumed to occur during the steady state, which has been well described in the literature. However, in many cases, imaging occurs during the transient stage, and the use of variable flip angles can improve signal and thus image quality. Here, we present the calculation of flip angles in transient balanced steady-state free precession to generate a predefined signal profile. The signal profile was iteratively optimized to maximize the integral of the signal versus time curve. The key contribution of this work is the formulation of the flip angle as a deterministic function of the preceding and desired magnetization. Catalyzation schemes, e.g., Kaiser-windowed ramp, can be combined with variable flip angles balanced steady-state free precession to reduce signal oscillations. A uniform signal profile was used as an example to demonstrate the variable flip angle algorithm. Accuracy of the algorithm and Bloch simulations were verified with MRI phantom acquisitions. Renal angiograms were acquired using an inflow-based balanced steady-state free precession MR angiography technique; improved small-vessel depiction was observed in volunteer examinations.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Angiography / instrumentation
  • Magnetic Resonance Angiography / methods*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity