Optical quality in eyes implanted with aspheric and spherical intraocular lenses assessed by NIDEK OPD-Scan: a randomized, bilateral, clinical trial

J Refract Surg. 2011 Apr;27(4):287-92. doi: 10.3928/1081597X-20100714-01. Epub 2010 Jul 15.

Abstract

Purpose: To determine whether implantation of an intraocular lens (IOL) with an aspheric surface (Akreos AO, Bausch & Lomb Inc) results in reduced ocular aberrations (spherical aberration) and improved Strehl ratio and modulation transfer function (MTF) after cataract surgery.

Methods: In an intraindividual, randomized, double-masked, prospective study of 50 eyes (25 patients) with bilateral cataract, an IOL with modified anterior and posterior surfaces (Akreos AO) was implanted in one eye and a biconvex IOL with spherical surfaces (Akreos Fit, Bausch & Lomb Inc) implanted in the fellow eye. Ocular aberrations, Strehl ratio, and MTF curve with 4.5-, 5.0-, and 6.0-mm pupils were measured with a NIDEK OPD-Scan dynamic retinoscopy aberrometer 3 months after surgery. Uncorrected and corrected distance visual acuity (UDVA and CDVA, respectively) were also measured.

Results: No statistically significant difference was noted between eyes in postoperative UDVA and CDVA at 1 month. At 3 months, the Akreos AO IOL group obtained statistically significant lower values of higher order and spherical aberrations with 4.5-, 5.0-, and 6.0-mm pupil diameters than the Akreos Fit IOL group (P<.05). The value of Strehl ratio was statistically significantly higher in eyes with the Akreos AO IOL for 4.5- and 6.0-mm pupils (P<.05). The MTF curve was better in the Akreos AO IOL group in 4.5-, 5.0-, and 6.0-mm pupils (P<.05).

Conclusions: The aspheric Akreos AO IOL induced significantly less spherical aberration than the Akreos Fit IOL for 4.5-, 5.0-, and 6.0-mm pupils. Modulation transfer function and Strehl ratio were also better in eyes implanted with the Akreos AO IOL than the Akreos Fit.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Aberrometry / instrumentation
  • Aberrometry / methods*
  • Aged
  • Corneal Wavefront Aberration / physiopathology
  • Corneal Wavefront Aberration / prevention & control*
  • Double-Blind Method
  • Female
  • Functional Laterality
  • Humans
  • Lens Implantation, Intraocular*
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Phacoemulsification*
  • Prospective Studies
  • Prosthesis Design
  • Pseudophakia / physiopathology
  • Pupil / physiology
  • Retinoscopy
  • Visual Acuity / physiology