The development of aortic insufficiency in left ventricular assist device-supported patients
- PMID: 20739615
- PMCID: PMC3089421
- DOI: 10.1161/CIRCHEARTFAILURE.109.917765
The development of aortic insufficiency in left ventricular assist device-supported patients
Abstract
Background: Aortic insufficiency (AI) following left ventricular assist device (LVAD) placement can affect device performance. The aim of this study was to examine AI development following LVAD implantation.
Methods and results: Echocardiograms (n=315) from 78 subjects undergoing HeartMate-XVE (n=25 [32%]) or HeartMate-II (n=53 [68%]) implantations from 2004 to 2008 were reviewed. Studies were obtained preoperatively and at 1, 3, 6, 12, 18, and 24 months after surgery. AI was graded on an interval scale (0=none, 0.5=trivial, 1=mild, 1.5=mild-moderate, 2=moderate, 2.5=moderate-severe, 3=severe), and the change in AI at follow-up was analyzed with significance tests. Kaplan-Meier estimates for freedom from moderate or worse AI at follow-up were generated. Mixed-model linear regression was used to identify correlates of AI progression during LVAD support. The median (25th, 75th percentile) duration of LVAD support was 239 (112, 455) days, and preoperative AI grade was 0.0 (0.0, 0.0). At 6 months, 89±4% of subjects (n=49 at risk) were free from moderate or worse AI, but this was reduced to 74±7% (n=29 at risk) and 49±13% (n=13 at risk) by 12 and 18 months, respectively. Correlates (slope±SE) of AI progression included female sex (0.002±0.001; P=0.01), smaller body surface area (-0.003±0.001 per m(2); P=0.0017), and HeartMate-II model type (0.002±0.001; P=0.039). Correlates (β±SE) of progressive AI on postoperative echocardiogram included increasing aortic sinus diameter (0.04±0.01 per mm; P=0.001), an aortic valve that remained closed (0.42±0.06; P<0.001) or only intermittently opened (0.34±0.09; P<0.001), and lower left ventricular diastolic (-0.002±0.0004 per cm(3); P<0.001) and systolic (-0.002±0.0004 per cm(3); P<0.001) volumes.
Conclusions: AI progresses over time in LVAD-supported patients. As we move toward an era of long-term cardiac support, more studies are needed to determine the clinical significance of these findings.
Figures
Comment in
-
Unforeseen consequences of therapy with continuous-flow pumps.Circ Heart Fail. 2010 Nov;3(6):647-9. doi: 10.1161/CIRCHEARTFAILURE.110.959023. Circ Heart Fail. 2010. PMID: 21081738 No abstract available.
Similar articles
-
Aortic insufficiency in continuous-flow left ventricular assist device support patients is common but does not impact long-term mortality.J Heart Lung Transplant. 2017 Jan;36(1):91-96. doi: 10.1016/j.healun.2016.07.018. Epub 2016 Jul 28. J Heart Lung Transplant. 2017. PMID: 27623098
-
Development of aortic insufficiency in patients supported with continuous flow left ventricular assist devices.ASAIO J. 2012 Jul-Aug;58(4):326-9. doi: 10.1097/MAT.0b013e318251cfff. ASAIO J. 2012. PMID: 22569164
-
The development of aortic insufficiency in continuous-flow left ventricular assist device-supported patients.Ann Thorac Surg. 2013 Feb;95(2):493-8. doi: 10.1016/j.athoracsur.2012.09.020. Epub 2012 Dec 13. Ann Thorac Surg. 2013. PMID: 23245444
-
De novo aortic insufficiency during long-term support on a left ventricular assist device: a systematic review and meta-analysis.ASAIO J. 2014 Mar-Apr;60(2):183-8. doi: 10.1097/MAT.0000000000000042. ASAIO J. 2014. PMID: 24399060 Review.
-
Percutaneous Transcatheter Interventions for Aortic Insufficiency in Continuous-Flow Left Ventricular Assist Device Patients: A Systematic Review and Meta-Analysis.ASAIO J. 2017 Mar/Apr;63(2):117-122. doi: 10.1097/MAT.0000000000000447. ASAIO J. 2017. PMID: 27676407 Review.
Cited by
-
Cellular and Molecular Mechanisms Activated by a Left Ventricular Assist Device.Int J Mol Sci. 2023 Dec 24;25(1):288. doi: 10.3390/ijms25010288. Int J Mol Sci. 2023. PMID: 38203459 Free PMC article. Review.
-
The Impact of Left Ventricular Assist Device Outflow Graft Positioning on Aortic Hemodynamics: Improving Flow Dynamics to Mitigate Aortic Insufficiency.Biomimetics (Basel). 2023 Oct 1;8(6):465. doi: 10.3390/biomimetics8060465. Biomimetics (Basel). 2023. PMID: 37887596 Free PMC article.
-
Transcatheter aortic valve implantation for aortic regurgitation in HeartMate II supported patient using Myval THV: a case report.Oxf Med Case Reports. 2023 Oct 23;2023(10):omad086. doi: 10.1093/omcr/omad086. eCollection 2023 Oct. Oxf Med Case Reports. 2023. PMID: 37881260 Free PMC article.
-
Hemodynamic Evaluation of Asynchronous Speed Modulation of a Continuous-Flow Left Ventricular Assist Device in an Acute-Myocardial Injury Sheep Model.Ann Biomed Eng. 2024 Feb;52(2):364-375. doi: 10.1007/s10439-023-03383-y. Epub 2023 Oct 18. Ann Biomed Eng. 2024. PMID: 37851145
-
Managing valvular pathology during LVAD implantation.Indian J Thorac Cardiovasc Surg. 2023 Jul;39(Suppl 1):101-113. doi: 10.1007/s12055-023-01567-8. Epub 2023 Jul 20. Indian J Thorac Cardiovasc Surg. 2023. PMID: 37525709
References
-
- Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH. Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med. 2007;357:885–96. - PubMed
-
- Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim DD, Tierney AR, Levitan RG, Watson JT, Meier P, Ronan NS, Shapiro PA, Lazar RM, Miller LW, Gupta L, Frazier OH, Desvigne-Nickens P, Oz MC, Poirier VL. Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med. 2001;345:1435–43. - PubMed
-
- Mudd JO, Cuda JD, Halushka M, Soderlund KA, Conte JV, Russell SD. Fusion of aortic valve commissures in patients supported by a continuous axial flow left ventricular assist device. J Heart Lung Transplant. 2008;27:1269–74. - PubMed
-
- Rose AG, Park SJ, Bank AJ, Miller LW. Partial aortic valve fusion induced by left ventricular assist device. Ann Thorac Surg. 2000;70:1270–4. - PubMed
-
- Rao V, Slater JP, Edwards NM, Naka Y, Oz MC. Surgical management of valvular disease in patients requiring left ventricular assist device support. Ann Thorac Surg. 2001;71:1448–53. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials