Comparison of a non-invasive arterial pulse contour technique and echo Doppler aorta velocity-time integral on stroke volume changes in optimization of cardiac resynchronization therapy

Europace. 2011 Jan;13(1):87-95. doi: 10.1093/europace/euq348. Epub 2010 Sep 29.

Abstract

Aims: We investigated the accuracy and feasibility of a non-invasive arterial pulse contour technique for continuous measurement of stroke volume (SV) in optimization of atrioventricular (AV) delay in cardiac resynchronization therapy (CRT), by comparing SV changes assessed by Nexfin CO-Trek® (Nexfin) and echo Doppler aortic velocity-time integral (VTIao). Furthermore, we investigated whether AV-delay optimization increases the effect of CRT when compared with a default AV delay (120 ms).

Methods and results: In 23 CRT patients, biventricular pacing (BiVP) was applied at various AV delays, while recording 10 beats preceding BiVP (baseline) and the first 10 BiVP beats, for both methods in parallel. Agreement between Nexfin and VTIao measurements was evaluated (Bland-Altman) on beat-to-beat changes in SV, as well as on effects of BiVP (averaged over 8 beats) at various AV delays. Individual optimal AV delays, for Nexfin (AVopt-n) and VTIao (AVopt-ao), were derived from the second-order polynomial fitted to the effect measurements of 20 patients. In 252 episodes assessed, the difference between measurements (= Nexfin - VTIao) was -0.6 ± 8.1% for beat-to-beat SV changes and -1.3 ± 7.3% for effects of BiVP. Optimal AV delays for Nexfin were well related to AVopt-ao (R(2) = 0.69). The effect (%) of BiVP at the optimal AV delay was significantly larger than at the default AV delay: median difference (range) being +6.3% (0.1-14.4%; P < 0.001) for VTIao and +4.7% (0.0-14.0%; P < 0.001) for Nexfin.

Conclusion: Individual AV optimization increases the effect of CRT. Nexfin is a promising tool in individual CRT optimization, as Nexfin agrees with VTIao on measuring beat-to-beat SV changes and on assessing relative effects of BiVP on SV at various AV delays.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aorta / physiopathology*
  • Atrioventricular Node / physiopathology
  • Blood Flow Velocity / physiology
  • Blood Pressure / physiology
  • Cardiac Resynchronization Therapy*
  • Defibrillators, Implantable
  • Echocardiography, Doppler / methods*
  • Feasibility Studies
  • Female
  • Heart Failure / physiopathology*
  • Heart Failure / therapy*
  • Heart Rate / physiology
  • Humans
  • Male
  • Pacemaker, Artificial
  • Pulse / instrumentation
  • Pulse / methods*
  • Reproducibility of Results
  • Stroke Volume / physiology*
  • Time Factors