Comparison of the erbium-yttrium aluminum garnet and carbon dioxide lasers for in vitro bone and cartilage ablation

Laryngoscope. 1990 Jan;100(1):14-7. doi: 10.1288/00005537-199001000-00004.

Abstract

The in vitro bone- and cartilage-ablation characteristics of the solid-state erbium:yttrium aluminum garnet laser were compared to those of the carbon dioxide laser. Ablations of fresh, frozen cadaver septal cartilage and maxillary sinus bone were performed using total energies between 1 and 6 J. Specimens were studied using hematoxylin and eosin stain and digitized, computer-assisted measurements of 35-mm photographs. Erbium-yttrium aluminum garnet-ablated bone averaged 5 microns of adjacent tissue thermal injury, compared with 67 microns with carbon dioxide-ablated bone. Erbium-yttrium aluminum garnet-ablated cartilage averaged 2 microns of adjacent tissue thermal injury, compared with 21 microns with the carbon dioxide-ablated cartilage. The tissue-ablation characteristics of the erbium-yttrium aluminum garnet laser are promising for future otolaryngologic applications.

Publication types

  • Comparative Study

MeSH terms

  • Aluminum
  • Bone and Bones / radiation effects
  • Bone and Bones / surgery*
  • Carbon Dioxide
  • Cartilage / radiation effects
  • Cartilage / surgery*
  • Erbium
  • Humans
  • Laser Therapy*
  • Yttrium

Substances

  • Carbon Dioxide
  • Yttrium
  • Erbium
  • Aluminum