Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure
- PMID: 21069287
- DOI: 10.1007/s00134-010-2076-1
Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure
Abstract
Purpose: Transcutaneous CO(2) monitors are widely used in neonatal ICUs. Until recently, these devices performed poorly in adults. Recent technical modifications have produced transcutaneous CO(2) monitors that have performed well in adults with chronic illnesses. We evaluated the accuracy of one of these devices, the TOSCA(®) 500, in adults admitted to an emergency department for acute respiratory failure.
Methods: We prospectively collected 29 pairs of simultaneous transcutaneous arterial CO(2) (PtcCO(2)) and arterial CO(2) (PaCO(2)) values in 21 consecutive adults with acute respiratory failure (acute heart failure, n = 6; COPD exacerbation, n = 8; acute pneumonia, n = 6; and pulmonary embolism, n = 1). Agreement between PaCO(2) and PtcCO(2) was evaluated using the Bland-Altman method.
Results: Mean arterial oxygen saturation was 90%, arterial oxygen tension ranged from 32 to 215 mmHg, and PaCO(2) ranged from 23 to 84 mmHg. The mean difference between PaCO(2) and PtcCO(2) was 0.1 mmHg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from -6 to 6.2 mmHg. None of the patients experienced adverse effects from heating of the device clipped to the earlobe.
Conclusion: PtcCO(2) showed good agreement with PaCO(2) in adults with acute respiratory failure.
Similar articles
-
Concordance between transcutaneous and arterial measurements of carbon dioxide in an ED.Am J Emerg Med. 2012 Nov;30(9):1872-6. doi: 10.1016/j.ajem.2012.03.033. Epub 2012 Jul 12. Am J Emerg Med. 2012. PMID: 22795407
-
Accuracy of three transcutaneous carbon dioxide monitors in critically ill children.Pediatr Pulmonol. 2010 May;45(5):481-6. doi: 10.1002/ppul.21203. Pediatr Pulmonol. 2010. PMID: 20425856
-
Accuracy of transcutaneous carbon dioxide monitoring in hypotensive patients.Emerg Med J. 2014 Apr;31(4):323-6. doi: 10.1136/emermed-2012-202228. Epub 2013 Feb 12. Emerg Med J. 2014. PMID: 23404804
-
Assessing PaCO2 in acute respiratory disease: accuracy of a transcutaneous carbon dioxide device.Intern Med J. 2011 Aug;41(8):630-3. doi: 10.1111/j.1445-5994.2010.02213.x. Epub 2010 Feb 26. Intern Med J. 2011. PMID: 20214687 Clinical Trial.
-
Concordance and limits between transcutaneous and arterial carbon dioxide pressure in emergency department patients with acute respiratory failure: a single-center prospective observational study.Scand J Trauma Resusc Emerg Med. 2015 May 17;23:40. doi: 10.1186/s13049-015-0120-4. Scand J Trauma Resusc Emerg Med. 2015. PMID: 25981461 Free PMC article.
Cited by
-
The utility of transcutaneous carbon dioxide measurements in the emergency department: A prospective cohort study.J Am Coll Emerg Physicians Open. 2021 Jul 17;2(4):e12513. doi: 10.1002/emp2.12513. eCollection 2021 Aug. J Am Coll Emerg Physicians Open. 2021. PMID: 34296208 Free PMC article.
-
Outcomes of Patients Treated with Prehospital Noninvasive Ventilation: Observational Retrospective Multicenter Study in the Northern French Alps.J Clin Med. 2021 Mar 25;10(7):1359. doi: 10.3390/jcm10071359. J Clin Med. 2021. PMID: 33806188 Free PMC article.
-
Regional capnometry to evaluate the adequacy of tissue perfusion.J Thorac Dis. 2019 Jul;11(Suppl 11):S1568-S1573. doi: 10.21037/jtd.2019.01.80. J Thorac Dis. 2019. PMID: 31388462 Free PMC article. Review.
-
How should we monitor patients with acute respiratory failure treated with noninvasive ventilation?Eur Respir Rev. 2018 Apr 13;27(148):170101. doi: 10.1183/16000617.0101-2017. Print 2018 Jun 30. Eur Respir Rev. 2018. PMID: 29653949 Free PMC article. Review.
-
Erratum: Concordance and limits between transcutaneous and arterial carbon dioxide pressure in emergency department patients with acute respiratory failure: a single-center, prospective, and observational study.Scand J Trauma Resusc Emerg Med. 2015 Oct 6;23:77. doi: 10.1186/s13049-015-0154-7. Scand J Trauma Resusc Emerg Med. 2015. PMID: 26572985 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources