Validation of the GENEA Accelerometer
- PMID: 21088628
- DOI: 10.1249/MSS.0b013e31820513be
Validation of the GENEA Accelerometer
Abstract
Purpose: The study aims were: 1) to assess the technical reliability and validity of the GENEA using a mechanical shaker; 2) to perform a GENEA value calibration to develop thresholds for sedentary and light-, moderate-, and vigorous-intensity physical activity; and 3) to compare the intensity classification of the GENEA with two widely used accelerometers.
Methods: A total of 47 GENEA accelerometers were attached to a shaker and vertically accelerated, generating 15 conditions of varying acceleration and/or frequency. Reliability was calculated using SD and intrainstrument and interinstrument coefficients of variation, whereas validity was assessed using Pearson correlation with the shaker acceleration as the criterion. Next, 60 adults wore a GENEA on each wrist and on the waist (alongside an ActiGraph and RT3 accelerometer) while completing 10-12 activity tasks. A portable metabolic gas analyzer provided the criterion measure of physical activity. Analyses involved the use of Pearson correlations to establish criterion and concurrent validity and receiver operating characteristic curves to establish intensity cut points.
Results: The GENEA demonstrated excellent technical reliability (CVintra = 1.4%, CVinter = 2.1%) and validity (r = 0.98, P < 0.001) using the mechanical shaker. The GENEA demonstrated excellent criterion validity using VO2 as the criterion (left wrist, r = 0.86; right wrist, r = 0.83; waist, r = 0.87), on par with the waist-worn ActiGraph and RT3. The GENEA demonstrated excellent concurrent validity compared with the ActiGraph (r = 0.92) and the RT3 (r = 0.97). The waist-worn GENEA had the greatest classification accuracy (area under the receiver operating characteristic curve (AUC) = 0.95), followed by the left (AUC = 0.93) and then the right wrist (AUC = 0.90). The accuracy of the waist-worn GENEA was virtually identical with that of the ActiGraph (AUC = 0.94) and RT3 (AUC = 0.95).
Conclusion: The GENEA is a reliable and valid measurement tool capable of classifying the intensity of physical activity in adults.
Similar articles
-
Calibration of the GENEA accelerometer for assessment of physical activity intensity in children.J Sci Med Sport. 2013 Mar;16(2):124-8. doi: 10.1016/j.jsams.2012.05.013. Epub 2012 Jul 6. J Sci Med Sport. 2013. PMID: 22770768
-
Physical activity classification using the GENEA wrist-worn accelerometer.Med Sci Sports Exerc. 2012 Apr;44(4):742-8. doi: 10.1249/MSS.0b013e31823bf95c. Med Sci Sports Exerc. 2012. PMID: 21988935
-
Establishing cut-points for physical activity classification using triaxial accelerometer in middle-aged recreational marathoners.PLoS One. 2018 Aug 29;13(8):e0202815. doi: 10.1371/journal.pone.0202815. eCollection 2018. PLoS One. 2018. PMID: 30157271 Free PMC article.
-
Classification accuracy of the wrist-worn gravity estimator of normal everyday activity accelerometer.Med Sci Sports Exerc. 2013 Oct;45(10):2012-9. doi: 10.1249/MSS.0b013e3182965249. Med Sci Sports Exerc. 2013. PMID: 23584403 Free PMC article.
-
Everything you wanted to know about selecting the "right" Actigraph accelerometer cut-points for youth, but…: a systematic review.J Sci Med Sport. 2012 Jul;15(4):311-21. doi: 10.1016/j.jsams.2011.12.001. Epub 2012 Feb 4. J Sci Med Sport. 2012. PMID: 22306372 Review.
Cited by
-
Assessing Neurobehavioral Alterations Among E-waste Recycling Workers in Hong Kong.Saf Health Work. 2024 Mar;15(1):9-16. doi: 10.1016/j.shaw.2023.12.004. Epub 2023 Dec 29. Saf Health Work. 2024. PMID: 38496288 Free PMC article.
-
Ecological Momentary Assessment to Explore Fatigue, Mood, and Physical Activity Levels in People Receiving Peritoneal Dialysis.Kidney Int Rep. 2023 Dec 30;9(3):601-610. doi: 10.1016/j.ekir.2023.12.024. eCollection 2024 Mar. Kidney Int Rep. 2023. PMID: 38481501 Free PMC article.
-
Comprehensive management of children and adolescents with type 1 diabetes mellitus through personalized physical exercise and education using an mHealth system: The Diactive-1 study protocol.Front Endocrinol (Lausanne). 2024 Feb 6;15:1354734. doi: 10.3389/fendo.2024.1354734. eCollection 2024. Front Endocrinol (Lausanne). 2024. PMID: 38379866 Free PMC article.
-
It needs more than a myocardial infarction to start exercising: the CoLaus|PsyCoLaus prospective study.BMC Cardiovasc Disord. 2024 Feb 12;24(1):102. doi: 10.1186/s12872-024-03755-9. BMC Cardiovasc Disord. 2024. PMID: 38347464 Free PMC article.
-
Study protocol for a 15-week randomised controlled trial assessing the independent effects of high-cholesterol and high-saturated fat diets on LDL cholesterol.BMJ Open. 2024 Jan 25;14(1):e081664. doi: 10.1136/bmjopen-2023-081664. BMJ Open. 2024. PMID: 38272555 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources