A meta-analysis and investigation for the source of bias of left ventricular volumes and function by three-dimensional echocardiography in comparison with magnetic resonance imaging

Am J Cardiol. 2011 Jan;107(1):126-38. doi: 10.1016/j.amjcard.2010.08.058.

Abstract

Determining accurate left ventricular (LV) function is clinically important. Three-dimensional echocardiography (3DE) achieves better estimation than 2-dimensional echocardiography. However, underestimation of LV volumes has often been reported without a systematic attempt to synthesize these data. This meta-analysis aimed to assess the bias of 3DE in evaluating LV volumes and ejection fraction (EF) and to investigate factors affecting that bias. Studies that compared LV volumes and/or EF between 3DE and magnetic resonance imaging were eligible. Meta-analysis of 95 studies including 3,055 subjects revealed significant underestimation of LV end-systolic volume (-4.7 ml, p <0.0001) and end-diastolic volume (-9.9 ml, p <0.0001), whereas measurement for EF revealed excellent accuracy (-0.13%, p = 0.41). Meta-regression analysis for factors of systematic bias in volumetry revealed that female gender and existence of cardiac disease were associated with more underestimation, whereas use of semiautomatic tracking and matrix-array transducers counteracted the underestimation. In conclusion, by meta-analysis synthesizing many small studies, we found underestimation of LV volumes and factors affecting the systematic bias by 3DE. These data provide a more detailed basis for analyzing and improving the accuracy of 3DE, an indispensable step toward further clinical application in LV assessment.

Publication types

  • Comparative Study
  • Meta-Analysis

MeSH terms

  • Bias
  • Echocardiography, Three-Dimensional
  • Female
  • Heart Ventricles / diagnostic imaging*
  • Humans
  • Magnetic Resonance Imaging
  • Male
  • Ventricular Function, Left*