Monte Carlo calculations of kQ, the beam quality conversion factor

Med Phys. 2010 Nov;37(11):5939-50. doi: 10.1118/1.3495537.

Abstract

Purpose: To use EGSnrc Monte Carlo simulations to directly calculate beam quality conversion factors, kQ, for 32 cylindrical ionization chambers over a range of beam qualities and to quantify the effect of systematic uncertainties on Monte Carlo calculations of kQ. These factors are required to use the TG-51 or TRS-398 clinical dosimetry protocols for calibrating external radiotherapy beams.

Methods: Ionization chambers are modeled either from blueprints or manufacturers' user's manuals. The dose-to-air in the chamber is calculated using the EGSnrc user-code egs_chamber using 11 different tabulated clinical photon spectra for the incident beams. The dose to a small volume of water is also calculated in the absence of the chamber at the midpoint of the chamber on its central axis. Using a simple equation, kQ is calculated from these quantities under the assumption that W/e is constant with energy and compared to TG-51 protocol and measured values.

Results: Polynomial fits to the Monte Carlo calculated kQ factors as a function of beam quality expressed as %dd(10)x and TPR10(20) are given for each ionization chamber. Differences are explained between Monte Carlo calculated values and values from the TG-51 protocol or calculated using the computer program used for TG-51 calculations. Systematic uncertainties in calculated kQ values are analyzed and amount to a maximum of one standard deviation uncertainty of 0.99% if one assumes that photon cross-section uncertainties are uncorrelated and 0.63% if they are assumed correlated. The largest components of the uncertainty are the constancy of W/e and the uncertainty in the cross-section for photons in water.

Conclusions: It is now possible to calculate kQ directly using Monte Carlo simulations. Monte Carlo calculations for most ionization chambers give results which are comparable to TG-51 values. Discrepancies can be explained using individual Monte Carlo calculations of various correction factors which are more accurate than previously used values. For small ionization chambers with central electrodes composed of high-Z materials, the effect of the central electrode is much larger than that for the aluminum electrodes in Farmer chambers.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorption
  • Air
  • Algorithms
  • Aluminum / chemistry
  • Calibration
  • Electrodes
  • Equipment Design
  • Humans
  • Models, Statistical
  • Monte Carlo Method
  • Photons
  • Radiometry / methods*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Water / chemistry

Substances

  • Water
  • Aluminum