Performance evaluation of a noninvasive hemoglobin monitoring device
- PMID: 21251729
- DOI: 10.1016/j.annemergmed.2010.11.032
Performance evaluation of a noninvasive hemoglobin monitoring device
Abstract
Study objective: Hemoglobin measurement is a routine procedure, and a noninvasive point-of-care device may increase the quality of care. The aim of the present study is to compare hemoglobin concentration values obtained with a portable totally noninvasive device, the Masimo Labs Radical-7 Pulse CO-Oximeter, with the results obtained by the ADVIA 2120 in the laboratory.
Methods: This was a prospective monocentric open trial enrolling patients consulting in the emergency department of a university hospital from June 16 to December 17, 2009. The main outcome measure was the agreement between both methods and evaluation of the percentage of potential decision error for transfusion.
Results: Samples from 300 consecutive patients were assessed. Hemoglobin concentration could not be obtained with the new device for 24 patients. In others, the mean bias, the lower and the upper limits of agreement between the 2 methods, was 1.8 g/dL (95% estimated confidence interval [CI] 1.5 to 2.1 g/dL), -3.3 g/dL (95% CI -3.8 to -2.8 g/dL), and 6.9 g/dL (95% CI 6.4 to 7.4 g/dL), respectively. The intraclass correlation coefficient was 0.53 (estimated 95% CI 0.10 to 0.74). The number of potential errors about transfusion decision was 38 (13% of patients). The peripheral oxygen saturation and the true value of hemoglobin concentration were independently associated with the bias.
Conclusion: Results from this widely available noninvasive point-of-care hemoglobin monitoring device were systematically biased and too unreliable to guide transfusion decisions.
Copyright © 2011 American College of Emergency Physicians. Published by Mosby, Inc. All rights reserved.
Comment in
-
Response to Gayat et al.Ann Emerg Med. 2011 Jul;58(1):106-7; author reply 107. doi: 10.1016/j.annemergmed.2011.02.028. Ann Emerg Med. 2011. PMID: 21689571 No abstract available.
Similar articles
-
Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients.Crit Care Med. 2011 Oct;39(10):2277-82. doi: 10.1097/CCM.0b013e3182227e2d. Crit Care Med. 2011. PMID: 21666449
-
Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds.Crit Care Med. 2012 Sep;40(9):2576-82. doi: 10.1097/CCM.0b013e318258fb4f. Crit Care Med. 2012. PMID: 22732278
-
Accuracy of noninvasive and invasive point-of-care total blood hemoglobin measurement in an outpatient setting.Postgrad Med. 2012 Jul;124(4):250-5. doi: 10.3810/pgm.2012.07.2584. Postgrad Med. 2012. PMID: 22913913
-
[Evaluation of circulatory state using pulse oximeter: 2. Measurement of total hemoglobin employing pulse CO-oximetry (Masimo Radical-7)].Masui. 2009 Jul;58(7):854-9. Masui. 2009. PMID: 19618826 Review. Japanese.
-
The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration.Anesth Analg. 2012 May;114(5):972-8. doi: 10.1213/ANE.0b013e318233041a. Epub 2011 Sep 29. Anesth Analg. 2012. PMID: 21965370 Review.
Cited by
-
Smartphone-based point-of-care anemia screening in rural Bihar in India.Commun Med (Lond). 2023 Mar 22;3(1):38. doi: 10.1038/s43856-023-00267-z. Commun Med (Lond). 2023. PMID: 36949164 Free PMC article.
-
Hemoglobin Determination Using Pulse Co-Oximetry and Reduced-Volume Blood Gas Analysis in the Critically Ill: A Prospective Cohort Study.Diagnostics (Basel). 2022 Nov 22;12(12):2908. doi: 10.3390/diagnostics12122908. Diagnostics (Basel). 2022. PMID: 36552914 Free PMC article.
-
Continuous hemoglobin measurement during frontal advancement operations can improve patient outcomes.J Clin Monit Comput. 2022 Dec;36(6):1689-1695. doi: 10.1007/s10877-022-00813-5. Epub 2022 Mar 7. J Clin Monit Comput. 2022. PMID: 35257263
-
Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.J Anesth. 2019 Jun;33(3):441-453. doi: 10.1007/s00540-019-02629-1. Epub 2019 Mar 20. J Anesth. 2019. PMID: 30895376
-
Non-invasive detection of severe neutropenia in chemotherapy patients by optical imaging of nailfold microcirculation.Sci Rep. 2018 Mar 28;8(1):5301. doi: 10.1038/s41598-018-23591-0. Sci Rep. 2018. PMID: 29593221 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources