Abstract

Cutaneous transilluminators are light-emitting devices used to localize blood vessels for various medical procedures. They are often used in populations that may be at increased risk for skin burns, such as neonates and the elderly. While there is a known potential for skin burns, little is known about the ophthalmic risk from the use of these devices. This paper will report on the laboratory evaluation of the potential ocular hazards from transilluminators (TIs). Our results indicate that transilluminators which incorporate white-light LEDs have emissions that have the potential for producing injury to the retina, especially in patients who may have a reduced aversion response.