In-vitro investigation of the hemodynamics of the Edwards Sapien transcatheter heart valve
- PMID: 21404898
In-vitro investigation of the hemodynamics of the Edwards Sapien transcatheter heart valve
Abstract
Background and aim of the study: The study aim was to evaluate the hemodynamic performance of the Sapien transcatheter heart valve (THV) in the ascending aorta, and its influence on the aortic wall in an in-vitro set-up.
Methods: A two-dimensional particle image velocimetry (2D-PIV) study was conducted to evaluate the hemodynamic performance of the Edwards Sapien THV in the aortic flow field (image rate 15 Hz). The prosthesis (diameter 23 mm) was placed inside a mock aorta under pulsatile flow conditions. The velocities, shear strength, vorticity and strain rate were obtained and calculated with a fixed frequency (70 Hz) at constant stroke volume (70 ml).
Results: The Sapien THV showed a jet flow-type profile with a maximum velocity of 0.87 +/- 0.16 m/s during peak flow phase (PFP). The jet flow was surrounded by ambilateral vortices with a higher percentage of counterclockwise than clockwise vorticity (335 +/- 66/s versus 277 +/- 44.1/s), analogous to the strain rate (261 +/- 55/s for elongation versus -168 +/- 25/s for contraction). The maximum shear strength was 26,284 +/- 11,550/s2, while the point-of-interest analysis revealed a higher velocity for the bottom aortic wall compared to the upper aortic wall (0.25 +/- 0.05 m/s versus 0.30 +/- 0.04 m/s; p = 0.014). All values were lower during the acceleration and deceleration phases compared to PFP.
Conclusion: The peak flow of the Sapien THV seems to be slightly higher than that of the native aortic valve, thus imitating near-physiological conditions. That the shear strength, vorticity and strain rate were high during peak flow phase, but low during other phases, might also have an influence on the aortic wall.
Similar articles
-
Effect of structural compliance on cavitation threshold measurement of mechanical heart valves.J Heart Valve Dis. 1994 Apr;3 Suppl 1:S77-83; discussion S83-4. J Heart Valve Dis. 1994. PMID: 8061872
-
Fluid-structure interaction of deformable aortic prostheses with a bileaflet mechanical valve.J Biomech. 2011 Jun 3;44(9):1684-90. doi: 10.1016/j.jbiomech.2011.03.036. Epub 2011 Apr 14. J Biomech. 2011. PMID: 21496823
-
In vitro pulsatile flow hemodynamics of five mechanical aortic heart valve prostheses.Eur J Cardiothorac Surg. 1992;6 Suppl 1:S113-23. Eur J Cardiothorac Surg. 1992. PMID: 1389270
-
Transcatheter pulmonary valve implantation using the Edwards SAPIEN transcatheter heart valve.Catheter Cardiovasc Interv. 2010 Feb 1;75(2):286-94. doi: 10.1002/ccd.22250. Catheter Cardiovasc Interv. 2010. PMID: 19924775
-
The utility of trans-catheter aortic valve replacement after commercialization: does the European experience provide a glimpse into the future use of this technology in the United States?Catheter Cardiovasc Interv. 2010 Mar 1;75(4):511-8. doi: 10.1002/ccd.22346. Catheter Cardiovasc Interv. 2010. PMID: 20088019 Review.
Cited by
-
Impact of Systemic Inflammatory Response Syndrome on Clinical, Echocardiographic, and Computed Tomographic Outcomes Among Patients Undergoing Transcatheter Aortic Valve Implantation.Front Cardiovasc Med. 2022 Feb 9;8:746774. doi: 10.3389/fcvm.2021.746774. eCollection 2021. Front Cardiovasc Med. 2022. PMID: 35224023 Free PMC article.
-
Does transcatheter aortic valve alignment matter?Open Heart. 2019 Nov 21;6(2):e001132. doi: 10.1136/openhrt-2019-001132. eCollection 2019. Open Heart. 2019. PMID: 31803486 Free PMC article.
-
Fluid-Structure Interaction Study of Transcatheter Aortic Valve Dynamics Using Smoothed Particle Hydrodynamics.Cardiovasc Eng Technol. 2016 Dec;7(4):374-388. doi: 10.1007/s13239-016-0285-7. Epub 2016 Nov 14. Cardiovasc Eng Technol. 2016. PMID: 27844463 Free PMC article.
-
The effects of positioning of transcatheter aortic valves on fluid dynamics of the aortic root.ASAIO J. 2014 Sep-Oct;60(5):545-552. doi: 10.1097/MAT.0000000000000107. ASAIO J. 2014. PMID: 25010918 Free PMC article.
-
Modeling single ventricle physiology: review of engineering tools to study first stage palliation of hypoplastic left heart syndrome.Front Pediatr. 2013 Oct 30;1:31. doi: 10.3389/fped.2013.00031. Front Pediatr. 2013. PMID: 24400277 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Research Materials