Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment

Radiother Oncol. 2011 Apr;99(1):86-9. doi: 10.1016/j.radonc.2011.01.024. Epub 2011 Mar 30.

Abstract

Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.

MeSH terms

  • Female
  • Head and Neck Neoplasms / radiotherapy
  • Humans
  • Male
  • Particle Accelerators
  • Prostatic Neoplasms / radiotherapy
  • Quality Assurance, Health Care
  • Radiometry
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / instrumentation*
  • Radiotherapy, Intensity-Modulated / instrumentation*
  • Spinal Neoplasms / radiotherapy