Dual-chamber pacemaker malfunction mimicking atrial capture by the ventricular electrode

Hellenic J Cardiol. 2011 Mar-Apr;52(2):171-6.

Abstract

A 78-year-old man came for a scheduled check up one month after the implantation of a permanent dualchamber pacemaker for symptomatic transient atrioventricular conduction disturbances (intermittent Mobitz II atrioventricular block). The patient's ECG indicated a loss of atrial capture by the atrial electrode with 1:1 atrioventricular conduction. A subsequent X-ray examination showed that both electrodes were well placed and in their correct sites. A detailed check of the pacemaker using the programmer, together with an echocardiographic examination, revealed the true nature of the malfunction: there was a significant delay between the atrial capture and atrial depolarisation and systole, as well as a loss of ventricular capture because of an acute increase in the threshold. Normal pacemaker function and pacing ECG were restored through modification of the pacemaker's functional parameters.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cardiac Pacing, Artificial / adverse effects*
  • Diagnosis, Differential
  • Equipment Failure*
  • Heart Block / diagnosis*
  • Heart Block / etiology*
  • Heart Block / therapy
  • Humans
  • Male
  • Pacemaker, Artificial / adverse effects*