Skip to main page content
Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
, 28 (6), 404-11

The Pneumatic Tourniquet: Mechanical, Ischaemia-Reperfusion and Systemic Effects

Affiliations
Review

The Pneumatic Tourniquet: Mechanical, Ischaemia-Reperfusion and Systemic Effects

Jean-Pierre Estebe et al. Eur J Anaesthesiol.

Abstract

The pneumatic tourniquet is frequently used for upper and lower limb surgery to reduce bleeding, improve visualisation of important structures and expedite surgical procedures. Despite advances in technology, localised tissue damage secondary to cuff compression, ischaemia-reperfusion injuries and systemic complications still occur. The combination of these problems may affect outcome and contribute to prolonged hospitalisation. Use of the correct pneumatic tourniquet cuff size and a patient-specific cuff pressure with careful control of the duration of inflation may help reduce the incidence of these injuries. The efficacy of ischaemic preconditioning or postconditioning, and experimental treatments such as free radical scavenging, and use of nitric oxide synthetase inhibitors on endothelial dysfunction, systemic neutrophil activation and coagulation reactions needs to be established.

Similar articles

See all similar articles

Cited by 21 PubMed Central articles

See all "Cited by" articles

LinkOut - more resources

Feedback