Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assays

Clin Chim Acta. 2011 Aug 17;412(17-18):1550-3. doi: 10.1016/j.cca.2011.04.034. Epub 2011 May 6.

Abstract

Background: Pre-analytical error accounts for major total laboratory errors. We assessed the impacts of hemolysis, icterus, and lipemia on laboratory tests on Roche Cobas 6000.

Methods: Various concentrations of hemoglobin, bilirubin, or Intralipid® were added into the plasma to simulate hemolytic, icteric, or lipemic samples. The analytes were then measured on Roche Cobas 6000 and the change of the analyte concentrations was determined.

Results: For most of the chemistry assays, our data were in a good agreement with Roche package inserts. However some assays had significant interference at lower index values while others were affected at higher index than the Roche package inserts indicated. In addition, we observed the positive interference by hemolysis on ALT, lipase, total protein, potassium, and iron. Negative interference was noted on calcium and CK. Most of the immunoassays were not affected by hemoglobin, bilirubin, and lipids although there were a few exceptions. Several therapeutic drugs were affected either positively or negatively by hemolysis, icterus, or lipemia to a certain extent.

Conclusions: We have demonstrated some test interferences which have not been reported previously on the Cobas 6000. The implementation of the cut-off indices on Cobas 6000 would provide more accurate test result reporting.

MeSH terms

  • Bilirubin / blood*
  • Clinical Chemistry Tests
  • Hemoglobins / analysis*
  • Hemolysis
  • Humans
  • Lipids / blood*

Substances

  • Hemoglobins
  • Lipids
  • Bilirubin