The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
- PMID: 2168451
- DOI: 10.1097/00004872-199007000-00004
The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
Abstract
With the increasing manufacture of expensive systems for the measurement of ambulatory blood pressure there is a need for potential purchasers to be able to satisfy themselves that the systems have been evaluated according to agreed criteria. The British Hypertension Society has, therefore, drawn up a protocol of requirements for the evaluation of these devices. This protocol incorporates many features of the American National Standard for Non-Automated Sphygmomanometers but includes many additional features, such as strict criteria for observer training, interdevice variability testing before and after a month of ambulatory use, and a new system of analysis which permits the test system to be graded. It is recommended that manufacturers of ambulatory blood pressure measuring devices should obtain an unbiased evaluation according to a recognized standard before a device is marketed.
Comment in
-
British hypertension protocol.J Hypertens. 1991 Jun;9(6):575-7. doi: 10.1097/00004872-199106000-00017. J Hypertens. 1991. PMID: 1653300 No abstract available.
-
The assessment of portable, non-invasive blood pressure recorders.J Hypertens. 1990 Jul;8(7):591-3. doi: 10.1097/00004872-199007000-00001. J Hypertens. 1990. PMID: 2168448 No abstract available.
Similar articles
-
Validation requirements for ambulatory blood pressure measuring systems.J Hypertens Suppl. 1991 Dec;9(8):S13-5. J Hypertens Suppl. 1991. PMID: 1795192
-
Evaluation of blood pressure measuring devices with special reference to ambulatory systems.J Hypertens Suppl. 1990 Dec;8(7):S133-9. J Hypertens Suppl. 1990. PMID: 2095381
-
Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometer.J Hypertens. 1990 Jul;8(7):599-605. doi: 10.1097/00004872-199007000-00003. J Hypertens. 1990. PMID: 2168450
-
Evaluation of blood pressure measuring devices.Clin Exp Hypertens. 1993 Nov;15(6):1087-97. doi: 10.3109/10641969309037096. Clin Exp Hypertens. 1993. PMID: 8268876 Review.
-
Factors influencing validation of ambulatory blood pressure measuring devices.J Hypertens. 1995 Nov;13(11):1235-40. J Hypertens. 1995. PMID: 8984119 Review.
Cited by
-
Blood Pressure Estimation Based on PPG and ECG Signals Using Knowledge Distillation.Cardiovasc Eng Technol. 2024 Feb;15(1):39-51. doi: 10.1007/s13239-023-00695-x. Epub 2024 Jan 8. Cardiovasc Eng Technol. 2024. PMID: 38191807
-
Sex-modified association between grip strength and mild cognitive impairment: a cross-sectional and follow-up study in rural China.BMC Geriatr. 2023 Nov 2;23(1):710. doi: 10.1186/s12877-023-04376-1. BMC Geriatr. 2023. PMID: 37919648 Free PMC article.
-
Non-invasive arterial blood pressure measurement and SpO2 estimation using PPG signal: a deep learning framework.BMC Med Inform Decis Mak. 2023 Jul 21;23(1):131. doi: 10.1186/s12911-023-02215-2. BMC Med Inform Decis Mak. 2023. PMID: 37480040 Free PMC article.
-
Association between handgrip strength asymmetry and cognitive function across ethnicity in rural China: a cross-sectional study.Front Aging Neurosci. 2023 May 19;15:1191197. doi: 10.3389/fnagi.2023.1191197. eCollection 2023. Front Aging Neurosci. 2023. PMID: 37273648 Free PMC article.
-
[A non-contact continuous blood pressure measurement method based on video stream].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2023 Apr 25;40(2):249-256. doi: 10.7507/1001-5515.202211071. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2023. PMID: 37139755 Free PMC article. Chinese.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical