Clinical outcomes using a new crossover balloon occlusion technique for percutaneous closure after transfemoral aortic valve implantation

JACC Cardiovasc Interv. 2011 Aug;4(8):861-7. doi: 10.1016/j.jcin.2011.05.019.

Abstract

Objectives: This study sought to evaluate the technical success and clinical outcomes of an adjunctive crossover balloon occlusion technique (CBOT) combined with the 10-F Prostar percutaneous closure device (PCD) on the incidence of vascular and bleeding complications in patients after transfemoral transcatheter aortic valve implantation (TAVI).

Background: Vascular closure following large-vessel access has most commonly been performed using a surgical cut-down and repair procedure.

Methods: Between November 2008 and September 2010, 58 consecutive patients with severe aortic stenosis underwent TAVI via a retrograde femoral artery approach using the Edwards-SAPIEN transcatheter valve. Among these patients, 56 were treated with a CBOT using the "pre-close" technique and the 10-F Prostar system. The technical success of this new CBOT and the 30-day frequency of clinical events, including all-cause mortality, major vascular complications, and major bleeding (defined according to a modified version of the Valve Academic Research Consortium criteria), were assessed.

Results: Successful closure was obtained in all but 3 patients (94.6%). The 30-day frequencies of mortality, major vascular complications, and major bleeding were 7.1%, 14.3%, and 5.4% respectively. No deaths were directly related to access site complications. Fourteen patients (25%) received at least 1 transfusion during the index hospitalization, of which 8 (57.1%) were not related to vascular complications. The mean and median hospital lengths of stay were 7.8 and 6.0 days.

Conclusions: This new percutaneous adjunctive CBOT combined with the Prostar PCD resulted in controlled, safe, and successful percutaneous closure in most patients after TAVI.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / therapy*
  • Balloon Occlusion / adverse effects
  • Balloon Occlusion / instrumentation
  • Balloon Occlusion / methods*
  • Balloon Occlusion / mortality
  • Blood Transfusion
  • Cardiac Catheterization / adverse effects*
  • Cardiac Catheterization / instrumentation
  • Cardiac Catheterization / methods
  • Cardiac Catheterization / mortality
  • Catheters
  • Female
  • Femoral Artery* / diagnostic imaging
  • Heart Valve Prosthesis
  • Heart Valve Prosthesis Implantation / adverse effects*
  • Heart Valve Prosthesis Implantation / instrumentation
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis Implantation / mortality
  • Hemorrhage / etiology
  • Hemorrhage / mortality
  • Hemorrhage / prevention & control*
  • Humans
  • Length of Stay
  • Male
  • New York City
  • Prospective Studies
  • Punctures
  • Radiography, Interventional
  • Time Factors
  • Treatment Outcome