A modified subcutaneous implantable cardioverter-defibrillator implant in a patient with a previous left ventricular epicardial defibrillation patch

Europace. 2012 Jan;14(1):149-50. doi: 10.1093/europace/eur262. Epub 2011 Aug 27.

Abstract

We describe a case of subcutaneous implantable cardioverter-defibrillator (ICD) implant in a patient with an existing epicardial defibrillation patch. Potential issues with shock vector shielding were overcome by a modification of the generator implant site and poor sensing were successfully managed by programming a sensing vector which excluded the generator.