Using FDA reports to inform a classification for health information technology safety problems
- PMID: 21903979
- PMCID: PMC3240763
- DOI: 10.1136/amiajnl-2011-000369
Using FDA reports to inform a classification for health information technology safety problems
Abstract
Objective: To expand an emerging classification for problems with health information technology (HIT) using reports submitted to the US Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database.
Design: HIT events submitted to MAUDE were retrieved using a standardized search strategy. Using an emerging classification with 32 categories of HIT problems, a subset of relevant events were iteratively analyzed to identify new categories. Two coders then independently classified the remaining events into one or more categories. Free-text descriptions were analyzed to identify the consequences of events.
Measurements: Descriptive statistics by number of reported problems per category and by consequence; inter-rater reliability analysis using the κ statistic for the major categories and consequences.
Results: A search of 899 768 reports from January 2008 to July 2010 yielded 1100 reports about HIT. After removing duplicate and unrelated reports, 678 reports describing 436 events remained. The authors identified four new categories to describe problems with software functionality, system configuration, interface with devices, and network configuration; the authors' classification with 32 categories of HIT problems was expanded by the addition of these four categories. Examination of the 436 events revealed 712 problems, 96% were machine-related, and 4% were problems at the human-computer interface. Almost half (46%) of the events related to hazardous circumstances. Of the 46 events (11%) associated with patient harm, four deaths were linked to HIT problems (0.9% of 436 events).
Conclusions: Only 0.1% of the MAUDE reports searched were related to HIT. Nevertheless, Food and Drug Administration reports did prove to be a useful new source of information about the nature of software problems and their safety implications with potential to inform strategies for safe design and implementation.
Conflict of interest statement
Figures
Similar articles
-
Exploring Health Information Technology Events from FDA MAUDE Database.Stud Health Technol Inform. 2018;250:187-191. Stud Health Technol Inform. 2018. PMID: 29857426
-
Patient safety problems associated with heathcare information technology: an analysis of adverse events reported to the US Food and Drug Administration.AMIA Annu Symp Proc. 2011;2011:853-7. Epub 2011 Oct 22. AMIA Annu Symp Proc. 2011. PMID: 22195143 Free PMC article.
-
Generating a Health Information Technology Event Database from FDA MAUDE Reports.Stud Health Technol Inform. 2019 Aug 21;264:883-887. doi: 10.3233/SHTI190350. Stud Health Technol Inform. 2019. PMID: 31438051
-
Understanding a decade of safety reporting for sacral neuromodulation in the Food and Drug Administration Manufacturer and User Facility Device Experience database.Neurourol Urodyn. 2023 Nov;42(8):1655-1667. doi: 10.1002/nau.25248. Epub 2023 Jul 27. Neurourol Urodyn. 2023. PMID: 37497812 Review.
-
Adverse Events Associated With Ossicular Prostheses: Utility of a Federal Database.Otol Neurotol. 2022 Feb 1;43(2):e229-e234. doi: 10.1097/MAO.0000000000003408. Otol Neurotol. 2022. PMID: 34889828 Review.
Cited by
-
Understanding prescribing errors for system optimisation: the technology-related error mechanism classification.BMJ Health Care Inform. 2024 Nov 2;31(1):e100974. doi: 10.1136/bmjhci-2023-100974. BMJ Health Care Inform. 2024. PMID: 39488434 Free PMC article.
-
More than algorithms: an analysis of safety events involving ML-enabled medical devices reported to the FDA.J Am Med Inform Assoc. 2023 Jun 20;30(7):1227-1236. doi: 10.1093/jamia/ocad065. J Am Med Inform Assoc. 2023. PMID: 37071804 Free PMC article.
-
Smart Glasses for Supporting Distributed Care Work: Systematic Review.JMIR Med Inform. 2023 Feb 28;11:e44161. doi: 10.2196/44161. JMIR Med Inform. 2023. PMID: 36853760 Free PMC article. Review.
-
Health technology, quality and safety in a learning health system.Healthc Manage Forum. 2023 Mar;36(2):79-85. doi: 10.1177/08404704221139383. Epub 2022 Dec 23. Healthc Manage Forum. 2023. PMID: 36562483 Free PMC article.
-
Evidence synthesis, digital scribes, and translational challenges for artificial intelligence in healthcare.Cell Rep Med. 2022 Dec 20;3(12):100860. doi: 10.1016/j.xcrm.2022.100860. Epub 2022 Dec 12. Cell Rep Med. 2022. PMID: 36513071 Free PMC article. Review.
References
-
- US Office of the National Coordinator for Health IT HIT Policy Committee, Adoption/Certification Workgroup meeting February 25, 2010. http://healthit.hhs.gov/ (accessed Apr 2010).
-
- US Joint Commission on Accreditation of Healthcare Organizations, 2008. http://www.jointcommission.org/SentinelEvents/SentinelEventAlert/sea_42.htm
-
- Coiera E, Westbrook J, Wyatt J. The safety and quality of decision support systems. Yearb Med Inform 2006;45(Suppl 1):20–5 - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical