Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure
- PMID: 21931078
- DOI: 10.1161/CIRCULATIONAHA.111.043042
Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure
Abstract
Background: Heart failure is associated with frequent hospitalizations, often resulting from volume overload. Measurement of intrathoracic impedance with an implanted device with an audible patient alert may detect increases in pulmonary fluid retention early. We hypothesized that early intervention could prevent hospitalizations and affect outcome.
Methods and results: We studied 335 patients with chronic heart failure who had undergone implantation of an implantable cardioverter-defibrillator alone (18%) or with cardiac resynchronization therapy (82%). All devices featured a monitoring tool to track changes in intrathoracic impedance (OptiVol) and other diagnostic parameters. Patients were randomized to have information available to physicians and patients as an audible alert in case of preset threshold crossings (access arm) or not (control arm). The primary end point was a composite of all-cause mortality and heart failure hospitalizations. During 14.9±5.4 months, this occurred in 48 patients (29%) in the access arm and in 33 patients (20%) in the control arm (P=0.063; hazard ratio, 1.52; 95% confidence interval, 0.97-2.37). This was due mainly to more heart failure hospitalizations (hazard ratio, 1.79; 95% confidence interval, 1.08-2.95; P=0.022), whereas the number of deaths was comparable (19 versus 15; P=0.54). The number of outpatient visits was higher in the access arm (250 versus 84; P<0.0001), with relatively more signs of heart failure among control patients during outpatient visits. Although the trial was terminated as a result of slow enrollment, a post hoc futility analysis indicated that a positive result would have been unlikely.
Conclusion: Use of an implantable diagnostic tool to measure intrathoracic impedance with an audible patient alert did not improve outcome and increased heart failure hospitalizations and outpatient visits in heart failure patients. Clinical Trial Registration-URL: http://www.clinicaltrials.gov. Unique identifier: NCT 00480077.
Trial registration: ClinicalTrials.gov NCT00480077.
Comment in
-
T.M.I. (Too much information)?Circulation. 2011 Oct 18;124(16):1697-9. doi: 10.1161/CIRCULATIONAHA.111.067819. Epub 2011 Sep 19. Circulation. 2011. PMID: 21931082 No abstract available.
Similar articles
-
Changes in intrathoracic impedance are associated with subsequent risk of hospitalizations for acute decompensated heart failure: clinical utility of implanted device monitoring without a patient alert.J Card Fail. 2009 Aug;15(6):475-81. doi: 10.1016/j.cardfail.2009.01.012. Epub 2009 Mar 17. J Card Fail. 2009. PMID: 19643357
-
Can monitoring of intrathoracic impedance reduce morbidity and mortality in patients with chronic heart failure? Rationale and design of the Diagnostic Outcome Trial in Heart Failure (DOT-HF).Eur J Heart Fail. 2008 Sep;10(9):907-16. doi: 10.1016/j.ejheart.2008.06.016. Epub 2008 Aug 19. Eur J Heart Fail. 2008. PMID: 18715826 Clinical Trial.
-
Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial.Eur Heart J. 2011 Sep;32(18):2266-73. doi: 10.1093/eurheartj/ehr050. Epub 2011 Feb 28. Eur Heart J. 2011. PMID: 21362703 Clinical Trial.
-
Fundamentals of intrathoracic impedance monitoring in heart failure.Am J Cardiol. 2007 May 21;99(10A):3G-10G. doi: 10.1016/j.amjcard.2007.02.009. Epub 2007 Mar 9. Am J Cardiol. 2007. PMID: 17512421 Review.
-
Integrating device-based monitoring into clinical practice: insights from a large heart failure clinic.Am J Cardiol. 2007 May 21;99(10A):17G-22G. doi: 10.1016/j.amjcard.2007.02.038. Epub 2007 Mar 9. Am J Cardiol. 2007. PMID: 17512418 Review.
Cited by
-
Adequacy of Ambulatory Hemodynamic Assessments for Reducing All-Cause Mortality in Individuals With Heart Failure.Int J Heart Fail. 2024 Oct 28;6(4):149-158. doi: 10.36628/ijhf.2024.0021. eCollection 2024 Oct. Int J Heart Fail. 2024. PMID: 39513021 Free PMC article. Review.
-
Biotronik Implantable Cardioverter Defibrillator for Measurement of Intrathoracic Impedance: A Report of Two Cases.Ann Noninvasive Electrocardiol. 2024 Nov;29(6):e70015. doi: 10.1111/anec.70015. Ann Noninvasive Electrocardiol. 2024. PMID: 39451048 Free PMC article.
-
Prediction of heart failure events based on physiologic sensor data in HINODE defibrillator patients.ESC Heart Fail. 2024 Oct;11(5):3322-3331. doi: 10.1002/ehf2.14890. Epub 2024 Jul 2. ESC Heart Fail. 2024. PMID: 38956896 Free PMC article.
-
Wearable Bioimpedance Monitoring: Viewpoint for Application in Chronic Conditions.JMIR Biomed Eng. 2021 May 11;6(2):e22911. doi: 10.2196/22911. JMIR Biomed Eng. 2021. PMID: 38907374 Free PMC article.
-
Efficacy of the Cardiac Implantable Electronic Device Multisensory Triage-HF Algorithm in Heart Failure Care: A Real-World Clinical Experience.Sensors (Basel). 2024 Jun 5;24(11):3664. doi: 10.3390/s24113664. Sensors (Basel). 2024. PMID: 38894453 Free PMC article.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical