Comparison of alternative interpretive criteria for the HIV-1 Western blot and results of the Multispot HIV-1/HIV-2 Rapid Test for classifying HIV-1 and HIV-2 infections

J Clin Virol. 2011 Dec:52 Suppl 1:S23-7. doi: 10.1016/j.jcv.2011.09.020. Epub 2011 Oct 12.

Abstract

Background: HIV-1 Western blot (WB) may be positive in specimens from persons with HIV-2 infection due to cross-reactive antibodies. HIV-1 and HIV-2 infections may be identified using assays designed to differentiate HIV-1 and HIV-2 antibody reactivity.

Objectives: To evaluate the ability of the current CDC WB criteria, alternative more stringent HIV-1 WB criteria (2 env plus one gag or pol band) and the Multispot HIV-1/HIV-2 Rapid Test to accurately differentiate HIV-1 and HIV-2 infections.

Study design: Two panels were used to determine the ability of each method to properly classify HIV-1 and HIV-2 infections: an HIV-2 panel (n=114) determined to be HIV-2 antibody-positive by both Multispot and by a validated HIV-2 WB, and 2135 HIV-1/HIV-2 immunoassay repeatedly reactive (IA-RR) specimens from the New York State Department of Health Laboratory (NYS).

Results: By CDC WB criteria, 53 (46.5%) HIV-2 panel specimens were HIV-1 WB positive, 60 (52.6%) were indeterminate, and 1 (0.9%) was negative; the alternative WB criteria re-classified 75.5% of the positives as indeterminate. Among 2135 NYS IA-RR specimens, the alternative WB criteria increased the proportion of indeterminates by 0.8%. Only 6 (0.3%) of the NYS specimens were determined to be HIV-2 infections; all 6 were classified either as HIV-1 positive or indeterminate by both WB criteria, but were classified as HIV-2 (n=4) or HIV-1/2 undifferentiated (n=2) by Multispot.

Conclusions: The alternative WB criteria classified most of the HIV-2 specimens that were HIV-1 positive by CDC criteria as indeterminate, but also slightly increased the proportion of HIV-1 specimens classified as indeterminate. The WB indeterminate specimens would require further testing or follow-up to resolve the infection status, whereas Multispot directly distinguished HIV-1 from HIV-2.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Blotting, Western*
  • Centers for Disease Control and Prevention, U.S. / standards
  • Cross Reactions
  • HIV Antibodies / immunology
  • HIV Infections / classification*
  • HIV Infections / diagnosis*
  • HIV Infections / immunology
  • HIV Infections / virology
  • HIV-1 / immunology
  • HIV-1 / pathogenicity
  • HIV-2 / immunology
  • HIV-2 / pathogenicity
  • Humans
  • Immunoassay / methods*
  • Reagent Kits, Diagnostic
  • Sensitivity and Specificity
  • United States
  • gag Gene Products, Human Immunodeficiency Virus / genetics
  • pol Gene Products, Human Immunodeficiency Virus / genetics

Substances

  • HIV Antibodies
  • Reagent Kits, Diagnostic
  • gag Gene Products, Human Immunodeficiency Virus
  • pol Gene Products, Human Immunodeficiency Virus