Aim: To quantify manual fraction of inspired oxygen (FiO(2)) adjustments performed by caregivers in extremely low birth weight (ELBW; ≤1000 g) infants, in relation to oxygen saturation (SpO(2)) and bedside care.
Methods: In a single-centre study, FiO(2) , SpO(2) and alarm limits of ELBW infants were collected for 3 days continuously, while caregivers were filmed. A descriptive analysis, focused on manual FiO(2) adjustments, was performed.
Results: Twelve ELWB infants were included. Total recording time was 726 h. FiO(2) was increased 851 times and decreased 1309 times; median (range) step size was 5% (1% to 65%) and -3% (-1% to -65%), respectively. Wide variation of FiO(2) adjustments for equal levels of SpO(2) was observed in all included infants. One hundred and twenty-six of 136 FiO(2) adjustments with a step size ≥15% and 111 of 171 desaturations <70% were associated with medical or nursing procedures. When FiO(2) was >21%, alarm limits for SpO(2) were set according to protocol (88-94%) in 64% of the time. Within these periods, SpO(2) was >94% for 30% and <88% for 16% of the time.
Conclusions: Manual FiO(2) adjustments varied widely in frequency and step size. Deep desaturations and large FiO(2) adjustments were associated with medical or nursing procedures. When large adjustments are really necessary, it will be challenging to implement them in an automatic adjustment device.
© 2011 The Author(s)/Acta Paediatrica © 2011 Foundation Acta Paediatrica.