Respiratory function monitor guidance of mask ventilation in the delivery room: a feasibility study

J Pediatr. 2012 Mar;160(3):377-381.e2. doi: 10.1016/j.jpeds.2011.09.017. Epub 2011 Nov 5.

Abstract

Objective: To investigate whether using a respiratory function monitor (RFM) during mask resuscitation of preterm infants reduces face mask leak and improves tidal volume (V(T)).

Study design: Infants receiving mask resuscitation were randomized to have the display of an RFM (airway pressure, flow, and V(T) waves) either visible or masked.

Result: Twenty-six infants had the RFM visible, and 23 had the RFM masked. The median mask leak was 37% (IQR, 21%-54%) in the visible RFM group and 54% (IQR, 37%-82%) in the masked RFM group (P = .01). Mask repositioning was done in 19 infants (73%) of the visible group and in 6 infants (26%) of the masked group (P = .001). The median expired V(T) was similar in the 2 groups. Oxygen was provided to 61% of the visible RFM group and 87% of the RFM masked group (P = .044). Continuous positive airway pressure use was greater in the visible RFM group (73% vs 43%; P = .035). Intubation in the delivery room was done in 21% of the visible group and in 57% of the masked group (P = .035).

Conclusion: Using an RFM was associated with significantly less mask leak, more mask adjustments, and a lower rate of excessive V(T).

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Delivery Rooms*
  • Feasibility Studies
  • Humans
  • Infant, Newborn
  • Infant, Premature*
  • Intubation, Intratracheal
  • Masks*
  • Monitoring, Physiologic*
  • Oxygen Inhalation Therapy
  • Positive-Pressure Respiration* / instrumentation
  • Respiration*
  • Resuscitation*
  • Tidal Volume