A novel voltammetric sensor for amoxicillin based on nickel-curcumin complex modified carbon paste electrode

Bioelectrochemistry. 2012 Jun:85:44-9. doi: 10.1016/j.bioelechem.2011.11.010. Epub 2011 Dec 7.

Abstract

The electrocatalytic oxidation of amoxicillin was investigated on a nickel-based (Ni(II)-curcumin) chemically modified electrode. This modified electrode was prepared by electropolymerization of complex (curcumin = 1,7-bis[4-hydroxyl-3-methoxyphenyl]-1,6-heptadiene-3,5-dione) in alkaline solution. For the first time, the catalytic oxidation of amoxicillin was demonstrated by cyclic voltammetry, chronoamperometry, chronocoulometry and amperometry methods at the surface of this modified carbon paste electrode. The obtained results showed that NiOOH acts as an electrocatalyst for oxidation of amoxicillin. This electrocatalytic oxidation exhibited a good linear response for amoxicillin concentration over the range of 8 × 10⁻⁶-1×10⁻⁴ M with a detection limit of 5 × 10⁻⁶ M. Therefore, this electrocatalytic method was used as a simple, selective and rapid method able to determine amoxicillin in pharmaceutical preparations and biological media.

MeSH terms

  • Amoxicillin / analysis*
  • Anti-Inflammatory Agents, Non-Steroidal
  • Carbon / chemistry*
  • Curcumin / chemistry*
  • Electrochemical Techniques / methods*
  • Electrodes
  • Limit of Detection
  • Nickel / chemistry*
  • Oxidation-Reduction

Substances

  • Anti-Inflammatory Agents, Non-Steroidal
  • Carbon
  • Nickel
  • Amoxicillin
  • Curcumin