Image-guided radiotherapy for left-sided breast cancer patients: geometrical uncertainty of the heart
- PMID: 22270162
- DOI: 10.1016/j.ijrobp.2011.08.024
Image-guided radiotherapy for left-sided breast cancer patients: geometrical uncertainty of the heart
Abstract
Purpose: To quantify the geometrical uncertainties for the heart during radiotherapy treatment of left-sided breast cancer patients and to determine and validate planning organ at risk volume (PRV) margins.
Methods and materials: Twenty-two patients treated in supine position in 28 fractions with regularly acquired cone-beam computed tomography (CBCT) scans for offline setup correction were included. Retrospectively, the CBCT scans were reconstructed into 10-phase respiration correlated four-dimensional scans. The heart was registered in each breathing phase to the planning CT scan to establish the respiratory heart motion during the CBCT scan (σ(resp)). The average of the respiratory motion was calculated as the heart displacement error for a fraction. Subsequently, the systematic (Σ), random (σ), and total random (σ(tot)=σ(2)+σ(resp)(2)) errors of the heart position were calculated. Based on the errors a PRV margin for the heart was calculated to ensure that the maximum heart dose (D(max)) is not underestimated in at least 90% of the cases (M(heart) = 1.3Σ-0.5σ(tot)). All analysis were performed in left-right (LR), craniocaudal (CC), and anteroposterior (AP) directions with respect to both online and offline bony anatomy setup corrections. The PRV margin was validated by accumulating the dose to the heart based on the heart registrations and comparing the planned PRV D(max) to the accumulated heart D(max).
Results: For online setup correction, the cardiac geometrical uncertainties and PRV margins were ∑ = 2.2/3.2/2.1 mm, σ = 2.1/2.9/1.4 mm, and M(heart) = 1.6/2.3/1.3 mm for LR/CC/AP, respectively. For offline setup correction these were ∑ = 2.4/3.7/2.2 mm, σ = 2.9/4.1/2.7 mm, and M(heart) = 1.6/2.1/1.4 mm. Cardiac motion induced by breathing was σ(resp) = 1.4/2.9/1.4 mm for LR/CC/AP. The PRV D(max) underestimated the accumulated heart D(max) for 9.1% patients using online and 13.6% patients using offline bony anatomy setup correction, which validated that PRV margin size was adequate.
Conclusion: Considerable cardiac position variability relative to the bony anatomy was observed in breast cancer patients. A PRV margin can be used during treatment planning to take these uncertainties into account.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
Estimation of heart-position variability in 3D-surface-image-guided deep-inspiration breath-hold radiation therapy for left-sided breast cancer.Radiother Oncol. 2013 Dec;109(3):442-7. doi: 10.1016/j.radonc.2013.09.017. Epub 2013 Oct 22. Radiother Oncol. 2013. PMID: 24161569
-
Image-guided radiotherapy for breast cancer patients: surgical clips as surrogate for breast excision cavity.Int J Radiat Oncol Biol Phys. 2011 Nov 1;81(3):e187-95. doi: 10.1016/j.ijrobp.2010.12.027. Epub 2011 Feb 23. Int J Radiat Oncol Biol Phys. 2011. PMID: 21345615
-
Breast patient setup error assessment: comparison of electronic portal image devices and cone-beam computed tomography matching results.Int J Radiat Oncol Biol Phys. 2010 Nov 15;78(4):1235-43. doi: 10.1016/j.ijrobp.2009.12.021. Epub 2010 May 14. Int J Radiat Oncol Biol Phys. 2010. PMID: 20472368
-
A review of setup error in supine breast radiotherapy using cone-beam computed tomography.Med Dosim. 2016 Autumn;41(3):225-9. doi: 10.1016/j.meddos.2016.05.001. Epub 2016 Jun 14. Med Dosim. 2016. PMID: 27311516 Review.
-
[The primary tumor target displacement and its impact during radiotherapy of esophageal cancer].Zhonghua Zhong Liu Za Zhi. 2013 Aug;35(8):561-5. Zhonghua Zhong Liu Za Zhi. 2013. PMID: 24314211 Review. Chinese. No abstract available.
Cited by
-
Analysis of heart displacement during thoracic radiotherapy based on electrocardiograph-gated 4-dimensional magnetic resonance imaging.J Thorac Dis. 2023 May 30;15(5):2754-2762. doi: 10.21037/jtd-23-447. Epub 2023 May 24. J Thorac Dis. 2023. PMID: 37324080 Free PMC article.
-
MRI-LINAC: A transformative technology in radiation oncology.Front Oncol. 2023 Jan 27;13:1117874. doi: 10.3389/fonc.2023.1117874. eCollection 2023. Front Oncol. 2023. PMID: 36776309 Free PMC article. Review.
-
The initial experience of MRI-guided precision prone breast irradiation with daily adaptive planning in treating early stage breast cancer patients.Front Oncol. 2022 Nov 23;12:1048512. doi: 10.3389/fonc.2022.1048512. eCollection 2022. Front Oncol. 2022. PMID: 36505797 Free PMC article.
-
Exposure of the heart and cardiac valves in women irradiated for breast cancer 1970-2009.Clin Transl Radiat Oncol. 2022 Jul 16;36:132-139. doi: 10.1016/j.ctro.2022.07.004. eCollection 2022 Sep. Clin Transl Radiat Oncol. 2022. PMID: 36034326 Free PMC article.
-
AAPM task group report 302: Surface-guided radiotherapy.Med Phys. 2022 Apr;49(4):e82-e112. doi: 10.1002/mp.15532. Epub 2022 Mar 15. Med Phys. 2022. PMID: 35179229 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical