Spine stereotactic body radiotherapy utilizing cone-beam CT image-guidance with a robotic couch: intrafraction motion analysis accounting for all six degrees of freedom

Int J Radiat Oncol Biol Phys. 2012 Mar 1;82(3):e555-62. doi: 10.1016/j.ijrobp.2011.06.1980.

Abstract

Purpose: To evaluate the residual setup error and intrafraction motion following kilovoltage cone-beam CT (CBCT) image guidance, for immobilized spine stereotactic body radiotherapy (SBRT) patients, with positioning corrected for in all six degrees of freedom.

Methods and materials: Analysis is based on 42 consecutive patients (48 thoracic and/or lumbar metastases) treated with a total of 106 fractions and 307 image registrations. Following initial setup, a CBCT was acquired for patient alignment and a pretreatment CBCT taken to verify shifts and determine the residual setup error, followed by a midtreatment and posttreatment CBCT image. For 13 single-fraction SBRT patients, two midtreatment CBCT images were obtained. Initially, a 1.5-mm and 1° tolerance was used to reposition the patient following couch shifts which was subsequently reduced to 1 mm and 1° degree after the first 10 patients.

Results: Small positioning errors after the initial CBCT setup were observed, with 90% occurring within 1 mm and 97% within 1°. In analyzing the impact of the time interval for verification imaging (10 ± 3 min) and subsequent image acquisitions (17 ± 4 min), the residual setup error was not significantly different (p > 0.05). A significant difference (p = 0.04) in the average three-dimensional intrafraction positional deviations favoring a more strict tolerance in translation (1 mm vs. 1.5 mm) was observed. The absolute intrafraction motion averaged over all patients and all directions along x, y, and z axis (± SD) were 0.7 ± 0.5 mm and 0.5 ± 0.4 mm for the 1.5 mm and 1 mm tolerance, respectively. Based on a 1-mm and 1° correction threshold, the target was localized to within 1.2 mm and 0.9° with 95% confidence.

Conclusion: Near-rigid body immobilization, intrafraction CBCT imaging approximately every 15-20 min, and strict repositioning thresholds in six degrees of freedom yields minimal intrafraction motion allowing for safe spine SBRT delivery.

MeSH terms

  • Cone-Beam Computed Tomography / methods*
  • Fiducial Markers
  • Humans
  • Immobilization / methods
  • Movement*
  • Organs at Risk / diagnostic imaging
  • Organs at Risk / radiation effects
  • Radiation Injuries / prevention & control
  • Radiosurgery / instrumentation
  • Radiosurgery / methods*
  • Radiotherapy Planning, Computer-Assisted / instrumentation
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy Setup Errors* / prevention & control
  • Radiotherapy, Computer-Assisted / instrumentation
  • Radiotherapy, Computer-Assisted / methods*
  • Robotics / instrumentation
  • Robotics / methods*
  • Spinal Cord / diagnostic imaging
  • Spinal Cord / radiation effects
  • Spinal Neoplasms / diagnostic imaging
  • Spinal Neoplasms / secondary
  • Spinal Neoplasms / surgery*