Risk assessment of electromagnetic fields exposure with metallic orthopedic implants: a cadaveric study

Orthop Traumatol Surg Res. 2012 Feb;98(1):90-6. doi: 10.1016/j.otsr.2011.08.012. Epub 2012 Jan 27.

Abstract

Metallic materials are well known to strongly interact with electromagnetic fields. While biological effects of such field have been extensively studied, only few works dealt with the interactions of electromagnetic waves with passive metallic device implanted in biological system. Hence only several numerical and phantom simulation studies were focusing on this aspect, whereas no in situ anatomic experiment has been previously performed. In this study the effect of electromagnetic waves on eight different orthopaedic medical devices (six plates from 55 to 318mm length, a total knee and a total hip prosthesis) were explored on six human cadavers. To mimic a random environmental exposure resulting from the most common frequencies band used in domestic environment and medical applications (TV and radio broadcasting, cell phone communication, MRI, diathermy treatment), a multifrequency generator emitting in VHF, UHF, GSM and GCS frequency bands was used. The different medical devices were exposed to an electromagnetic field at 50W/m(2) and 100W/m(2). After 6min exposure, the temperature was measured on three points close to each medical device, and the induced currents were estimated. No significant temperature increase (<0.2°C) was finally detected; beside, a slight induced tension (up to 1.1V) was recorded but would appear too low to induce any biological side effect.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged, 80 and over
  • Cadaver
  • Electromagnetic Fields / adverse effects*
  • Environmental Exposure / adverse effects*
  • Female
  • Humans
  • Male
  • Materials Testing / methods*
  • Models, Anatomic*
  • Orthopedic Fixation Devices*
  • Prostheses and Implants*
  • Radiation Dosage
  • Risk Assessment / methods*
  • Temperature