Ultrasound-guided fine-needle aspiration biopsy of clinically suspicious thyroid nodules with an automatic aspirator: a novel technique
- PMID: 22524469
- DOI: 10.1089/thy.2011.0182
Ultrasound-guided fine-needle aspiration biopsy of clinically suspicious thyroid nodules with an automatic aspirator: a novel technique
Abstract
Background: Fine-needle aspiration biopsy (FNAB) is a simple technique for the investigation of suspicious thyroid nodules. However, low success rates are reported in the literature. The aim of this prospective study was to compare the clinical performance and impact of an automatic aspirator, referred to here as Aspirator 3, to those of the manual technique for the FNAB of clinically suspicious thyroid nodules.
Methods: One hundred nine consecutive patients with 121 clinically suspicious thyroid nodules underwent a biopsy twice of the same site with the clinically approved Aspirator 3 and with the manual technique. The number of follicular cell formations and the total number of follicular cells in the aspirate were counted using the ThinPrep® method.
Results: With the Aspirator 3, the total number and the mean number of extracted cell formations were significantly higher than the values achieved with the manual technique (total: 3222 vs. 1951, p=0.02; mean: 27 vs. 16). The total number of cells that were biopsied was also higher when the Aspirator 3 was utilized (47,480 vs. 23,080, p=0.005). Overall, the Aspirator 3 was superior in 65 biopsies, and the manual technique was superior in 39 biopsies.
Conclusions: In terms of cell formations and the total number of cells aspirated, the Aspirator 3 was superior to the manual technique. Further, the Aspirator 3 was more convenient to use and had a greater precision in needle guidance.
Similar articles
-
[Optimization of the fine needle aspiration biopsy (FNAB) of thyroid nodules- automatic aspirator versus manual technique].Nuklearmedizin. 2011;50(5):189-94. doi: 10.3413/nukmed-0387-11-02. Epub 2011 Jul 5. Nuklearmedizin. 2011. PMID: 21727989 German.
-
Success rate of repeated fine needle aspiration biopsy of clinically suspicious thyroid nodules.Nuklearmedizin. 2012;51(4):116-8. doi: 10.3413/Nukmed-0472-12-01. Epub 2012 May 3. Nuklearmedizin. 2012. PMID: 22552740
-
Effects of nodule characteristics on sampling number and duration of thyroid fine-needle aspiration biopsy: size does not matter, but cystic degeneration ratio does.Acta Radiol. 2017 Mar;58(3):286-291. doi: 10.1177/0284185116649797. Epub 2016 Jul 20. Acta Radiol. 2017. PMID: 27235454
-
[Diagnostic work-up of euthyroid nodules: which nodules should undergo fine-needle aspiration biopsy? Relevance of ultrasound].Dtsch Med Wochenschr. 2009 Dec;134(49):2498-503. doi: 10.1055/s-0029-1243052. Epub 2009 Nov 25. Dtsch Med Wochenschr. 2009. PMID: 19941231 Review. German.
-
Statement and Recommendations on Interventional Ultrasound as a Thyroid Diagnostic and Treatment Procedure.Ultrasound Med Biol. 2018 Jan;44(1):14-36. doi: 10.1016/j.ultrasmedbio.2017.08.1889. Epub 2017 Nov 7. Ultrasound Med Biol. 2018. PMID: 29126752 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources