MRI compatibility and safety of cochlear implants

Zhong Nan Da Xue Xue Bao Yi Xue Ban. 2012 Mar;37(3):311-5. doi: 10.3969/j.issn.1672-7347.2012.03.018.

Abstract

The increasing magnetic strengths used for magnetic resonance imaging (MRI) combined with a greater number of MRI patients with cochlear implants (CIs) raises possible safety issues. This paper reviews the interactions between MRI and CI, and assesses MRI compatibility and safety of CIs at 0.2, 0.3, 1.0, 1.5 and 3.0 tesla (T). In a word, CIs represent a significantcontraindication to MRI. Only after careful evaluation of the individual risk and careful preparations can an MRI be safely performed.

Publication types

  • Review

MeSH terms

  • Cochlear Implants* / adverse effects
  • Contraindications
  • Equipment Failure Analysis
  • Humans
  • Magnetic Resonance Imaging / adverse effects*
  • Magnetic Resonance Imaging / instrumentation