Multi-center experience of 164 consecutive Hemodialysis Reliable Outflow [HeRO] graft implants for hemodialysis treatment

Eur J Vasc Endovasc Surg. 2012 Jul;44(1):93-9. doi: 10.1016/j.ejvs.2012.04.011. Epub 2012 May 12.

Abstract

Objective: To report a multi-center experience with the novel Hemodialysis Reliable Outflow (HeRO) vascular access graft.

Materials and methods: Four centers conducted a retrospective review of end stage renal disease patients who received the HeRO device from implant to last available follow-up. Data is available on 164 patients with an accumulated 2092.1 HeRO implant months.

Results: At 6 months, HeRO primary and secondary patency is 60% and 90.8%, respectively and at 12 months, 48.8% and 90.8%, respectively. At 24 months, HeRO had a primary patency of 42.9% and secondary patency was 86.7%. Interventions to maintain or re-establish patency have been required in 71.3% of patients (117/164) resulting in an intervention rate of 1.5/year. Access related infections have been reported in 4.3% patients resulting in a rate of 0.14/1000 implant days.

Conclusions: In our experience the HeRO device has performed comparably to standard AVGs and has proven superior to TDCs in terms of patency, intervention, and infection rates when compared to the peer-reviewed literature. As an alternative to catheter dependence as a means for hemodialysis access, this graft could reduce the morbidity and mortality associated with TDCs and have a profound impact on the costs associated with catheter related infections and interventions.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Blood Vessel Prosthesis*
  • Catheters, Indwelling*
  • Equipment Failure Analysis / methods*
  • Female
  • Follow-Up Studies
  • Humans
  • Kidney Failure, Chronic / therapy*
  • Male
  • Middle Aged
  • Prosthesis Design
  • Renal Dialysis / instrumentation*
  • Retrospective Studies
  • Young Adult