DNA hybridization detection in microfluidic devices can reduce sample volumes, processing times, and can be integrated with other measurements. However, as device footprints decrease and their complexity increase, the signal-to-noise ratio in these systems also decreases and the sensitivity is thereby compromised. Device miniaturization produces distinct properties and phenomena with greater influence at the micro-scale than at the macro-scale. Here, a diffusion-restriction model was applied to a miniaturized biochip nanovolume reactor to accurately characterize DNA hybridization events that contribute to shifts in both charge transfer resistance and diffusional resistance. These effects are shown to play a significant role in electrochemical impedance spectroscopy (EIS) analyses at these length scales. Our highly functional microfluidic biosensor enables the detection of ssDNA targets selectively, with a calculated detection limit of 3.8 nM, and cross-reactivity of 13% following 20 min incubation with the target. This new biosensing approach can be further modeled and tested elucidating diffusion behavior in miniaturized devices and improving the performance of biosensors.
Copyright © 2012 Elsevier B.V. All rights reserved.