Quantitative measurement of hard x-ray spectra for high intensity laser produced plasma

Rev Sci Instrum. 2012 May;83(5):053502. doi: 10.1063/1.4717677.

Abstract

X-ray line spectra ranging from 17 to 77 keV were quantitatively measured with a Laue spectrometer, composed of a cylindrically curved crystal and a detector. Either a visible CCD detector coupled with a CsI phosphor screen or an imaging plate can be chosen, depending on the signal intensities and exposure times. The absolute sensitivity of the spectrometer system was calibrated using pre-characterized laser-produced x-ray sources and radioisotopes. The integrated reflectivity for the crystal is in good agreement with predictions by an open code for x-ray diffraction. The energy transfer efficiency from incident laser beams to hot electrons, as the energy transfer agency for specific x-ray line emissions, is derived as a consequence of this work.