Adequate disinfection of a split-septum needleless intravascular connector with a 5-second alcohol scrub
- PMID: 22669226
- DOI: 10.1086/666337
Adequate disinfection of a split-septum needleless intravascular connector with a 5-second alcohol scrub
Abstract
Objective: Define optimum vascular catheter connector valve disinfection practices under laboratory and clinical conditions.
Design: Prospective observational clinical survey and laboratory assessment of disinfection procedures.
Setting: All adult inpatients at an academic healthcare center.
Methods: In the clinical setting, contamination of needleless connectors was assessed in 6 weekly prevalence surveys in which the connector valves from central venous catheters (CVCs) in situ were cultured by pressing the connector diaphragm to an agar plate. Before culture, valves were disinfected by scrubbing the diaphragm with a 70% isopropyl alcohol pledget for 0, 5, 10, 15, or 30 seconds. In the laboratory, the diaphragms on 150 unused sterile connector valves were inoculated with 10(3), 10(5), or 10(8) colony-forming units of Staphylococcus epidermidis and allowed to dry. After disinfection of the diaphragms by scrubbing with a 70% isopropyl alcohol pledget for 0, 5, 10, 15, or 30 seconds, the valves were sampled by pressing the diaphragm to an agar plate.
Results: In the clinical setting, 363 connector valves from patients with CVCs were sampled, and 66.7% of nondisinfected valves revealed bacterial contamination. After 5-second disinfection with an alcohol pledget, only 1 (1.4%) of 71 yielded microbial growth (P < .005). In the laboratory, at the 10(3) and 10(5) inoculum, all connector valves yielded sterile cultures when scrubbed for 5 or more seconds (P < .001). At the 10(8) inoculum, 2 (20%) of 10 connector valves yielded minimal growth of S. epidermidis.
Conclusions: A 5-second scrub with a 70% isopropyl alcohol pledget yields adequate disinfection of a split-septum intravascular catheter connector valve under clinical and laboratory conditions.
Similar articles
-
Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap.Infect Control Hosp Epidemiol. 2006 Jan;27(1):23-7. doi: 10.1086/500280. Epub 2006 Jan 6. Infect Control Hosp Epidemiol. 2006. PMID: 16418982
-
Preventing central venous catheter-associated bloodstream infections: development of an antiseptic barrier cap for needleless connectors.Am J Infect Control. 2008 Dec;36(10):S174.e1-5. doi: 10.1016/j.ajic.2008.10.006. Am J Infect Control. 2008. PMID: 19084153
-
The colonization rate of needleless connector and the impact of disinfection for 15 s on colonization: A prospective pre- and post-intervention study.J Vasc Access. 2019 Nov;20(6):604-607. doi: 10.1177/1129729819826036. Epub 2019 Feb 5. J Vasc Access. 2019. PMID: 30722715
-
The effects of needleless connectors on catheter-related bloodstream infections.Am J Infect Control. 2011 May;39(4):277-83. doi: 10.1016/j.ajic.2010.07.011. Epub 2011 Jan 22. Am J Infect Control. 2011. PMID: 21256629 Review.
-
Challenges in reducing the risk of infection when accessing vascular catheters.J Hosp Infect. 2021 Jul;113:130-144. doi: 10.1016/j.jhin.2021.03.005. Epub 2021 Mar 11. J Hosp Infect. 2021. PMID: 33713758 Review.
Cited by
-
Bacterial transfer and biofilm formation in needleless connectors in a clinically simulated in vitro catheter model.Infect Control Hosp Epidemiol. 2023 Nov;44(11):1760-1768. doi: 10.1017/ice.2023.60. Epub 2023 Apr 24. Infect Control Hosp Epidemiol. 2023. PMID: 37088696 Free PMC article.
-
Minimising Blood Stream Infection: Developing New Materials for Intravascular Catheters.Medicines (Basel). 2020 Aug 26;7(9):49. doi: 10.3390/medicines7090049. Medicines (Basel). 2020. PMID: 32858838 Free PMC article. Review.
-
Needleless connectors: the vascular access catheter's microbial gatekeeper.J Infect Prev. 2016 Sep;17(5):234-240. doi: 10.1177/1757177416657164. Epub 2016 Jul 25. J Infect Prev. 2016. PMID: 28989484 Free PMC article.
-
Prevention of Device-Related Healthcare-Associated Infections.F1000Res. 2016 Jan 14;5:F1000 Faculty Rev-65. doi: 10.12688/f1000research.7493.1. eCollection 2016. F1000Res. 2016. PMID: 26918162 Free PMC article. Review.
-
In vitro study on the disinfectability of two split-septum needle-free connection devices using different disinfection procedures.GMS Hyg Infect Control. 2015 Dec 9;10:Doc17. doi: 10.3205/dgkh000260. eCollection 2015. GMS Hyg Infect Control. 2015. PMID: 26693394 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources